background image

7.

Use the 200 mL measuring glass to collect the sample of the three calibration cycles.

8.

Enter the total volume of the three cycles. Push 

NEXT

.

The sampler shows the message "LOWER VOLUME CALIBRATED"

The sampler is calibrated and prepared for operation.

Replace the pump tube

W A R N I N G

 

Pinch hazard. Parts that move can pinch and cause injury. Do not touch moving parts.

W A R N I N G

 

Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the instrument or
equipment malfunction. The replacement parts in this section are approved by the manufacturer.

C A U T I O N

 

Pinch hazard. Remove power from the instrument before maintenance or service activities are done.

Examine and clean the pump tubing and rollers at regulars intervals. Replace the pump tubing at regular intervals
and when damage is seen.

Items to collect:

Silicone tubing for peristaltic pump, item no. BM901062 

Refer to the illustrated steps replace the pump tube.

Note: Make sure that the hose and hose connectors are dry.

2