4
MA391572_13.5
[Original Instructions]
7/01/2013
Water Connections
1. Connect one end of the supplied ¼-inch water hose to the water outlet fitting located on the
back of the HandiMet 2.
•
Measure and cut the appropriate length of hose to install the recommended Buehler
Water Filtration Kit (60-9091).
2. Measure and cut the appropriate length of hose to extend to an external water source.
3. Connect the other end of the water hose to an external water source equipped with a
separate shut-off valve.
Figure 1 HandiMet 2 Drain and Water Connections
NOTICE
The main water supply must be shut off at night or when the machine is left unattended.
Drain Connections
1. Place the HandiMet 2 on a level surface for proper drainage.
2. Connect one end of the supplied drain hose to the drain outlet fitting located on the back of
the
HandiMet 2.
3. Extend the remaining length of the drain hose to a drain, sink, recirculating system, or other
available catch container.
NOTICE
Place the drain hose at an incline to allow for proper drainage and to prevent the buildup
of debris residue in the hose.