3
MA391572_13.5
[Original Instructions]
7/01/2013
Unpacking
The HandiMet 2 has been carefully packaged for protection during transit from the factory to your
location.
If any components are missing or damaged, save the packing list and materials and advise Buehler and
the carrier of any discrepancies.
Carefully unpack and check that the following items have been received:
•
240 grit CARBIMET™
abrasive paper roll
•
320 grit CARBIMET™
abrasive paper roll
•
400 grit CARBIMET™
abrasive paper roll
•
600 grit CARBIMET™
abrasive paper roll
•
Drain hose (48 inches)
•
Water hose (48 inches)
•
2 Paper support rods
CAUTION
!
Equipment Damage. The HandiMet 2 weighs approximately 33 lbs (15 kg). Carefully lift the unit
from the shipping carton.
Follow approved procedures and safety practices when lifting and installing this machine.
Improper lifting can result in machine damage.
Plumbing
Select a location for your HandiMet 2 that provides an adequate working space, a power source, water
connections, and a drain.
When connecting water to the HandiMet 2 observe local water regulations. Use a Buehler connection kit if
using a pressure reliever and a check valve from the water source.
NOTICE
Use of the Buehler Water Filtration Kit (60-9091) is highly recommended. Not using the Water
Filtration Kit can cause damage to the machine and any damages will not be covered under
warranty.
Plumbing installation and machine maintenance should be performed by qualified personnel.
The recommended water pressure for the HandiMet 2 is 40 to 100 psi.