6 Operation
39
V-850 / V-855 Operation Manual, Version E
Use the selection knob on the vacuum controller to move the “cursor” from one character to the other
and press
Enter
each time you want to take a character over into the field at the top. Press
Save
to
save the solvent under the entered name.
To delete a self-defined solvent, in the Edit solvent screen go to Delete solvent, select the desired
solvent from the drop-down list, press
OK
and confirm with
Yes
.
To determine whether the predefined solvents should be shown or hidden in the solvent library, go to
Predefined solvents and press the Forward button. The following screen appears:
You can now hide or show individual solvents by going to Hide or Show, selecting the corresponding
solvents from the drop-down list and pressing
OK
. To hide or show all predefined solvents, go to Hide
all or Show all and confirm by pressing
OK
.
Calibr . AutoDest probe
By means of this submenu you can calibrate your AutoDest probe, if applicable. For this purpose,
follow the instructions on the screen.
NOTE
During the calibration procedure the system carries out a plausibility check on the measured values
and displays an error message when the confirmed values are implausible.
For a description on how to calibrate the pressure sensor, see the separate section 6.8.