Operation Manual Dehumidifier S-396 Nano
19/24
7.4
Cleaning and servicing the warning and directive symbols
u
Check that the warning symbols on the instrument are legible.
u
If they are dirty, clean them.
7.5
Cleaning the ventilation slots
u
Remove dust and foreign objects from the ventilation slots using compressed air
or a vacuum cleaner.
7.6
Cleaning the instrument
u
Disconnect the power cable.
u
Remove the process gas tubes.
u
Open the cap nuts.
u
Remove the bottle.
u
Put the condensate drain tubes in a bucket
(min. 5 L)
u
Seal the condensate drain tubes with a hose
clamp.
u
Fill in about 5 liter cleaning liquid (e.g. ethanol)
into the process gas tube connected to the
connection marked IN.
u
Wait 5 minutes.
NOTICE! Check the filling level of the bucket dur-
ing the draining process. If necessary empty it.
u
Drain the cleaning liquid through the
condensate drain tubes.
u
Repeat the cleaning procedure until the
washing solution is free from particles or
chemicals