6 Operation
34
R II Operation Manual, Version E
6 .3
Selecting the distillation conditions
To achieve optimal distillation conditions, the distillation energy supplied by the heating bath must be
removed by the condenser.
To ensure this, operate the instrument according to the following rule of thumb:
Cooling water: max . 20 °C
Vapor: 40 °C
Bath: 60 °C
How are these conditions achieved?:
• Set the bath temperature to 60 °C.
• Set the cooling water temperature not higher than 20 °C.
• Allow cooling water to flow through the condenser at approximately 40 – 50 L/h.
• Define the operating vacuum in such a way, that the boiling point of the solvent is 40 °C. The
corresponding pressure can be seen from the Solvent Table in chapter 3.
Advantages associated with bath temperatures of 60 °C:
• The evaporating flask can be replaced without risk of burns.
• The evaporation rate of the water from the heating bath is low (low energy loss).
• The heating bath energy is used at a good degree of efficiency.
This rule can also be applied to lower bath temperatures, e.g.:
Cooling water: 0 °C
Vapor: 20 °C
Bath: 40 °C
Δ T
2
(min. 20 °C)
Δ T
1
(min. 20 °C)
Fig. 6.2: 20–40–60 °C rule
Содержание Rotavapor R II
Страница 1: ...Operation Manual Rotavapor R II 093102 en...
Страница 2: ......
Страница 6: ...1 About this manual 6 R II Operation Manual Version E...
Страница 20: ...3 Technical data 20 R II Operation Manual Version E...
Страница 32: ...5 Putting into operation 32 R II Operation Manual Version E...
Страница 42: ...8 Troubleshooting 42 R II Operation Manual Version E...
Страница 54: ...10 Spare parts 54 R II Operation Manual Version E...
Страница 56: ...11 Declarations and requirements 56 R II Operation Manual Version E 11 2 Declaration of conformity...
Страница 57: ......