7 Maintenance and repairs
75
NIRFlex N-500
Operation Manual, Version G
Resetting the laser lifetime counter
After LASER unit replacement the following message appears:
Message concerning required instrument setup
• Click the OK icon
to confirm the message.
• Now click the Service icon
(Operator).
The following dialog opens:
Assembly setup dialog
• Click the icon next to the Laser Module row
. The following dialog appears:
Exchange sequence dialog for the laser
• Activate the watch icon next to the line
Laser hours of operation
.
The following message box is displayed:
Message
• Confirm the message by clicking the OK icon
.