11 | Cleaning and servicing
BÜCHI Labortechnik AG
74/88
Operation Manual FatExtractor E-500
Component
Action
Frequency
Heating plate
u
Wipe down the heating plate with a
damp cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
u
If cracked or broken, contact BUCHI
customer service
Weekly
Hoses and hose
connections
u
Check the hoses and hose connec-
tions for defects (cracks, brittle ar-
eas).
u
If defective, replace defect hoses.
Weekly
Housing
u
Wipe down the housing with a damp
cloth.
u
If heavily soiled, use ethanol or a mild
detergent.
Weekly
Warning symbols
u
Check that the warning symbols on
the instrument are legible.
u
If they are dirty, clean them.
Weekly
Protection shields
NOTICE! Do not use Acetone
u
Wipe down the protection shields with
a damp cloth.
u
Make sure that they click into place. If
not, replace the gliding elements. See
Replacing the gliding elements
u
If defective or damaged, replace
them.
Weekly
Reflector foil for an-
alyte protection
NOTICE! Do not use any other deter-
gent
u
Switch the On/Off master switch to
Off.
u
Make sure that there are no extrac-
tion beakers in the instrument.
u
Check the reflector foil for dirt.
u
If dirty, wipe down the reflector foil
with a damp cloth.
u
Switch the
On/Off
master switch to
On.
Weekly
Interface
u
Wipe down the display with a damp
cloth.
Monthly
11.2
Removing the tank bottle
u
Switch the
On/Off
master switch to Off.