Footswitch and electrogoniometer kits
34
Finally, another 5 seconds’ pause ❷ allows moving the sensing block at an angle of
-90° with
respect to the reference block
❸
. Once again, the sensing block must be held in the selected
positio
n for 10 seconds ❹ (
Figure 7.20
At the end of
Channel 2’s
calibration process, once again, the calculated angle offset and sensitivity
are displayed. To save the calibration press on
Ok
).
At this stage, the calibration of the EGN probe has been completed. The described procedure must
be reiterated for each electrogoniometer which needs to be calibrated (
).
❸
❹
❶
❷