6.  USER SELECT MODE 

This mode allows you to save the resistance of a user wire (See NOTE below) for additional 
measurements of unknown lengths of the same gauge wire. In addition, it enables you to 
accurately measure the length of standard gauge wires. In this mode, you can measure the 
length of any metal wire, the resistance of which can be measured, as well as Copper or 
Aluminum wires. 

NOTE: The sample length of user wires must be 5m in METER or 20ft in FEET 
mode. 

 
How to save the resistance of a user wire 
 

Note: You will need a 5M (or 20ft in feet mode) sample length of the wire you are 
programming into the meter.    This will measure the resistance of that wire and 
store a value to correctly measure longer lengths. 

1.  Turn on the meter and perform the calibration procedure first. 
2.  Allow the meter to attain the same temperature as the user wire. 
 

 

WARNING: MAKE SURE WIRE UNDER TEST IS NOT 

ENERGIZED. NEVER APPLY VOLTAGE TO INPUTS.

    

 

3.  Strip the insulation on each end of the user wire. 

NOTE: Make sure both ends of the wire under test are clean and the conductor is fully 
exposed. The insulation must be stripped away so the Kelvin clips can be good connected 
across the diameter of the wire. 

4.  Using the selector on the mete, turn to the decided memory location in the User Select 

range. The meter has 8 internal memory locations from 1 to 8, and the selected memory 
location number will illuminate at the top left corner of the display. For example, “user 1” 
on the display means memory location 1 is being selected 
If the selected memory location is occupied, the display shows “0000”. 
If the selected memory location is empty, the display shows “no”. 

5.  Press the “FT” or “M” button to select the measurement unit. 
6.  Connect a Kelvin clip to one end of the user wire and the other Kelvin clip to the next end 

of the wire. 

7.  Press “MEM” button to enter the Memory function mode. The “MEM” enunciator will 

illuminate at the top of the display. 

8.  Press “CAL” button to store the resistance of the user wire to the selected memory 

location. 

9.  If there have any stored resistance value in the selected memory location, the meter will 

display “ SurE”. In this case, press “CAL” button again to store the new resistance value. 

10. Press “MEM” button or turn the selector to any position will exit Memory function mode. 
11. Disconnect the test leads from the meter after saving the resistance value. 
Tips: User can write down the user wire data at the bottom of the rubber holster. There have a 
Listing label for user to keep memory location record. 

 

 

Содержание BS33

Страница 1: ...User s Manual MODEL Cable Length Meter BS33 C Copyright BSTCaltek 2011 REV2 1 MAR 11...

Страница 2: ...mend the specifications of the mentioned products without notice 2 Index 1 INTRODUCTION 3 2 FEATURES 3 3 SPECIFICATIONS 4 4 CALIBRATION PROCEDURE 6 5 MEASURING LENGTH OF WIRE 7 6 USER SELECT MODE 8 7...

Страница 3: ...ntories of wire 2 FEATURES 4 digits display with enunciators Measures in FEET ft or METERS m Direct readout for both COPPER Cu or ALUMINUM Al wire User Select Mode with 8 memory locations Optimised fo...

Страница 4: ...wing table to determine the minimum and maximum length that can be measured for specific gauge wires GAUGE mm2 Min M Max M Min Ft Max Ft 240 29 5 30 00k 97 0 100 0k 185 26 1 30 00k 86 0 100 0k 150 24...

Страница 5: ...eading 1m or 3ft 100m or 300ft below 18 and above 23 2 of reading 100m or 300ft below 18 and above 23 Use the User Select mode for better measurement accuracies Resistance 0 to 10 0 5 of reading 3 LSD...

Страница 6: ...r of input jacks This is very important to ensure that the meter works correctly If black and red test leads are mixed the meter will give false readings See figure1 4 Figure1 3 Connect the Kelvin cli...

Страница 7: ...good connected across the diameter of the wire 4 Using the selector on the BS33 turn to the size of wire under test 5 Press the COPPER button if copper wire is being tested The Cu enunciator will illu...

Страница 8: ...d The insulation must be stripped away so the Kelvin clips can be good connected across the diameter of the wire 4 Using the selector on the mete turn to the decided memory location in the User Select...

Страница 9: ...lect the decided memory location 5 Press the FT or M button to select the measurement unit 6 Connect a Kelvin clip to one of the wire under test and the other Kelvin clip to the next end of the wire 7...

Страница 10: ...OPERATIONAL HINTS Never apply voltage to the inputs Temperature affects readings Allow the meter to attain the ambient temperature as the wire under test The sample length of user wires for the User...

Страница 11: ...sed by shipment using NOT the original packaging will be charged in any case to the consignor The manufacturer will not be responsible for any damage to persons or things The warranty does not apply t...

Страница 12: ...nch the outward transport is payable by the customer The delivery must be agreed in advance with consignee For delivery indicate the reasons for returning it By means of an enclosed note as clear as p...

Отзывы: