16
Read carefully the instructions for use of the control solutions that you are using.
It is possible to check the proper working of MULTICARE
IN
device by using the MULTICARE
IN
control
solutions. The control solutions can be purchased as a separate accessory through specialized retailers.
The control solution is applied on the strips just like blood. When the test is performed with the control
solutions, the result must fall within the range printed on the strips vial label that you are using. Write
down the date of the first opening on the vial of the control solution. The control solutions must be used
within 90 days from the first opening. Do not use control solutions after the expiration date.
You should perform a control solutions test when:
• Suspecting a possible defect of the MULTICARE
IN
device.
• Suspecting the possibility of a malfunction or wrong measurement.
The result obtained with the control solutions must be deleted by the memory of the device in order to
not interfere with the average calculation of the results, in particular with the average of the glycemia.
MULTICAREIN CONTROL SOLUTIONS
CONTROL SOLUTIONS