8
Diode test
Defaults at
. Press
SELECT button momentarily 3 times to select
Diode test
function. Normal forward voltage drop (forward biased) for a good silicon diode is
between 0.400V to 0.900V. A reading higher than that indicates a leaky diode
(defective). A zero reading indicates a shorted diode (defective). An OL indicates an
open diode (defective). Reverse the test leads connections (reverse biased) across the
diode. The digital display shows OL if the diode is good. Any other readings indicate
the diode is resistive or shorted (defective).
Temperature (Model
906s & 907s
only)
Defaults at
o
C. Press
SELECT button momentarily to select
o
F function.
Note: Be sure to insert the banana plug type-K temperature bead probe Bkp60 with
correct
polarities. You can also use a plug adapter Bkb32 (Optional purchase)
with banana pins to type-K socket to adapt other standard type-K mini plug
temperature probes.