![Brymen BM905 Скачать руководство пользователя страница 14](http://html.mh-extra.com/html/brymen/bm905/bm905_user-manual_2823001014.webp)
13
Battery and Fuse replacement
Battery use:
Standard 1.5V AAA Size (NEDA 24A or IEC LR03) battery X 2
Fuses use:
Fuse (FS1) for
AmA current input:
0.4A
/1000V ac & dc, IR 30kA, F fuse, or better; Dimension: 6 x 32 mm
Fuse (FS2) for A current input:
11A/1000V ac & dc, IR 20kA, F fuse, or better; Dimension: 10 x 38 mm
Battery replacement for models with
battery access door:
Loosen the 2 screws from the
battery access door of the case
bottom. Lift the battery access door
and thus the battery compartment
up. Replace the battery. Re-fasten
the screws.
Fuse replacement (and also Battery
replacement for splash proof version
without battery access door):
Loosen the 4 screws from the case
bottom. Lift the end of the case
bottom nearest the input jacks until it
unsnaps from the case top. Replace
the blown fuse(s) and/or the battery.
Replace the case bottom, and
ensure that all the gaskets are
properly seated and the two snaps
on the case top (near the LCD side)
are engaged. Re-fasten the screws.
Содержание BM905