24
Ohms
RANGE
Accuracy
600.0
0.9% + 5d
6.000k
, 60.00k
, 600.0k
0.9% + 2d
6.000M
1.2% + 3d
60.00M
3.0% + 6d
Open Circuit Voltage: < 1.5VDC typical
Audible Continuity Tester
Audible threshold: between 20
and 200
Response time < 30ms
Diode Tester
Range
Accuracy
Test Current (Typical)
Open Circuit Voltage
2.000V
1.5% + 4d
0.5mA
< 2.8 VDC
Capacitance (Model 878 only)
RANGE
Accuracy
1)
3.000
F
2)
, 30.00
F, 300.0
F, 3000
F
1.5% + 5d
30.00mF
10% + 5d
1)
Accuracies with film capacitor or better
2)
Readings not available below 200nF
DCmV (Model 878 only)
RANGE
Accuracy
60.00mV,
0.5% + 3d
600.0mV,
0.1% + 3d
Input impedance: 10M
, 140pF nominal
ACmV (Model 878 only)
RANGE
Accuracy
50Hz ~ 60Hz
1% + 3d
60Hz ~ 3kHz
2% + 3d
60.00mV, 600.0mV
3kHz ~ 5kHz
3% + 5d
Input impedance: 10M
, 140pF nominal
Содержание BM876
Страница 1: ...USER S MANUAL BM878 BM877 BM876 Insulation Tester Combination Multimeter ...
Страница 15: ...14 ...