15
COMPARE mode (Models 877 & 876 only)
This mode uses a preset insulation resistance threshold value for PASS/FAIL
measuring comparison. The LCD annuciator
turns on when the
reading is
higher than the selected threshold value. On the contrary, the LCD annuciator
turns on and the meter chirps when the
reading is below such value.
To enable, press the
COMPARE button momentarily. The LCD annuciator turns on.
Press the
COMPARE button momentarily again to select the preset threshold values in
sequence. Last threshold values selected for each test voltage range will be saved
separately as power up default for repeat measurement convenience. Press the
COMPARE button for 1 second or more to exit.
NOTE. Selectable threshold values for the test voltage ranges are 100kΩ/ 200kΩ/
500kΩ/ 1MΩ/ 2MΩ/ 5MΩ/ 10MΩ/ 20MΩ/ 50MΩ for 50V & up. Additional 100MΩ for
100V & up, 200MΩ for 250V & up and 500MΩ for 500V & 1000V ranges.
Содержание BM876
Страница 1: ...USER S MANUAL BM878 BM877 BM876 Insulation Tester Combination Multimeter ...
Страница 15: ...14 ...