22
General Specification
Display:
4-5/6 digits 60,000 counts.
Polarity:
Automatic
Update Rate:
4-5/6 digits: Max 5 per second nominal
31 Segment Bar-graph: 50 per second max
Operating Temperature:
-20
C to 55
C continuous operating (except on
A
function, see
Electrical Specifications below for more details)
Relative Humidity:
Maximum relative humidity 80% for temperature up to 31
C decreasing
linearly to 50% relative humidity at 55
C
Pollution degree:
2
Storage Temperature:
-20
C to 60
C, < 80% R.H. (with battery removed)
Altitude:
Operating below 2000m
Temperature Coefficient:
nominal 0.10 x (specified accuracy)/
C @ (-20
C
~
18
C or
28
C
~
55
C), or otherwise specified
Sensing:
AC and AC+DC True RMS
Safety:
Double insulation per IEC/UL/EN 61010-1 Ed. 3.0, IEC/UL/EN 61010-2-030 Ed. 1.0,
IEC/UL/EN 61010-2-033 Ed. 1.0, IEC/UL/EN 61010-031 Ed. 2.0 and the corresponding
CAN/CSA-C22.2 regulations to Measurement Categories III 1000V AC & DC and Category
IV 600V AC & DC
Overload Protections:
A & mA: 0.4A/1000V DC/AC, IR 30kA or better, F fuse
A: 11A/1000V DC/AC, IR 20kA or better, F fuse
V: 1100V DC/AC rms
mV,
& Others: 1000 V DC/AC rms
Transient protection:
8kV (1.2/50
s surge)
E.M.C.:
Meets EN61326-1:2013
Power Supply:
1.5V AAA Alkaline battery x 3
Power Consumption:
10mA typical for AC & AC+DC Voltage/Current functions; 8mA
typical for other functions
Low Battery:
Below approx. 3.7V
APO Timing:
Idle for 15 minutes
APO Consumption:
15
A typical.
Dimension:
L193mm X W89mm X H51mm
Weight:
635 gm
Accessories:
Test lead pair, User's manual, Bkp60 banana plug K-type thermocouple x 1
(Models
789 & 786
only)
Optional Accessories:
BKB32 banana plug to type-K socket plug adaptor (Models
789 &
786
only), BMH-02 magnetic hanger strap
Special Features:
AutoHold; VFD; BeepLit
TM
Continuity; Record MAX, MIN, & AVG