Introduction
Z33119_2_003
15
3.2.1
General Information
Data
Value
Weight
~ 30 kg
Length
48 cm
Width
35.1 cm
Height
42.5 cm
Transfer line length
2, 3, 5 or 8 m
Sound pressure level
< 60 dB(A)
Electrical Information
Type 1
220 V – 240 V ~ 50 Hz
Max. 2.1 A
Type 2
115 V ~ 60 Hz
Max. 4.7 A
Type 3
200 V – 220 V ~ 50 Hz
200 V – 220 V ~ 60 Hz
Max. 2.2 A
Max. 2.4 A
Power Requirement
600 VA
Power Consumption
280 W
Cool down time
~15 min
Circuit protection
2 Fuses 5 x 20 mm 10 AT
Gas Supply
Data
Value
Min. gas temperature (end of
transfer line)
-40 °C @ 0 – 3000 l/h & 17…25 °C ambient
temperature
Gas dew point temperature
≤ -50 °C @ 1 bar
Flow rate
0…3000 Nl/h
Operating Environment
Data
Value
Ambient operating temperature
range
Min 17 °C, max 32 °C, 17…25 °C, specifications
fulfilled
For the appropriate temperature see also the Bruker site planning guides on the BASH CD
(Bruker Advanced Service Handbook).
Содержание SmartCooler BCU I
Страница 1: ...SmartCooler BCU I User Manual Version 003 Innovation with Integrity NMR...
Страница 12: ...Safety 12 Z33119_2_003...
Страница 18: ...Introduction 18 Z33119_2_003...
Страница 28: ...Preparation 28 Z33119_2_003...
Страница 30: ...Installation 30 Z33119_2_003...
Страница 39: ...Troubleshooting Z33119_2_003 39 9 Troubleshooting Please refer to the table on the following page...
Страница 42: ...Troubleshooting 42 Z33119_1_003...
Страница 44: ...Replacement of Parts 44 Z33119_2_003...
Страница 48: ...Contact 48 Z33119_2_003...
Страница 49: ...Z33119_2_003 49...
Страница 50: ...Bruker Corporation info bruker com www bruker com Order No Z33119...