Installation
Z33146_1_001
25 / 56
The installation of the iProbe HR-Liquids is complete and the iProbe HR-Liquids can be
detected by the software with the command
edprobe
.
6.2.2
Installation of iProbe HR-MAS
NOTICE
In the iProbe HR-MAS the bearing gas has two functions: supporting the rotor for spinning
and regulating its temperature. There is no dedicated VT gas flow unless the probe is
specified as DVT. In this last case, the setup of the VT GAS is analogous to the iProbe HR-
liquids.
The following instructions are intended for non-DVT probes only.
Prerequisite: The quick-lock is mounted to the
lower part of the shim system according to the
Quick-Lock Mounting Instruction
(Z166745).
• Take the iProbe HR-MAS out of the
transport box and remove the protection
cap.
• Check that the quick lock lever is in the
unlock position.
• Slide the iProbe HR-MAS into the magnet.
Mount the iProbe so that the clip holders
(1) on the iProbe body fit exactly the holes
for the clip holders in the quick-lock so that
the iProbe is in the correct orientation.
• Lock the iProbe HR-MAS in the position
using the quick-lock lever.
Connections at the rear side:
• Remove the white plastic cover and
connect the Johnston connector of the VT-
gas hose or the BCU to the rear side of the
iProbe HR-MAS.
• Mount the support clamp (Z53660) on the
magnet as shown in the photo. Position
and support the transfer line using the
Velcro tape of the support clamp.
Содержание NMR iProbe HR-Liquids
Страница 1: ...iProbes HR Liquids HR MAS User Manual Version 001 Innovation with Integrity NMR ...
Страница 44: ...Maintenance 44 56 Z33146_1_001 ...
Страница 46: ...Troubleshooting 46 56 Z33146_1_001 ...
Страница 48: ...Replacement of Parts 48 56 Z33146_1_001 ...
Страница 54: ...Contact 54 56 Z33146_1_001 ...
Страница 55: ...Z33146_1_001 55 56 ...
Страница 56: ...Bruker Corporation info bruker com www bruker com Order No Z33146 ...