• Once the vial has been removed, use the steel rod or pestle to crush the remaining glass
within the temperature controller. Repeatedly press down until the glass has been reduced
to fine granules. Use the rod/pestle in the surrounding openings in the event any of the
breakage spilled into them.
Note:
If the module is cooling, set the temperature to 65°C to liquefy any fat which has
accumulated at the bottom of the block.
Figure 8.10: Steel Rod
Figure 8.11: Breaking the Remaining Glass into Fine Granules with the Steel Rod
• Use the shop vacuum with the attachment to extract the glass particles. Also clean all of
the surrounding holes.