Storage
If it is necessary to store the magnet system before installation comply with the following
conditions:
•
Store the transport box in a closed, dry and dust-free room.
•
Store the transport box upright.
•
Do not tilt the transport box.
•
Do not unpack the transport box.
•
Prevent from mechanical vibrations.
•
Storage temperature: 5 – 40 °C.
•
Storage humidity: less than 50 % @ 23 °C.
Figure 3.2:
Transportation by forklift / pallet jack
1.
Check the route of transport for the minimum
height and width.
2.
Check sufficient floor capacity on the route of
transport. In case of doubt ask a stress
analyst.
3.
Check sufficient carrying capacity while using
an elevator.
4.
Position the forks between the bars of the
transport box as shown in the figure.
5.
Make sure the forks project out of the back of
the transport box as shown in the figure below.
6.
Now lift the fork and move the transport box to
the site.
Содержание EPR 12T Cryogen Free
Страница 1: ...Version Innovation with Integrity EPR 12T Cryogen Free 01 User Manual EPR Spectroscopy ...
Страница 6: ...6 ZTKS0213 Z31939 Rev 01 Contact ...
Страница 25: ...25 ZTKS0213 Z31939 Rev 01 4 Assembling Approved persons Bruker Service only ...
Страница 26: ...26 ZTKS0213 Z31939 Rev 01 Assembling ...
Страница 48: ...48 ZTKS0213 Z31939 Rev 01 Operation ...
Страница 54: ...54 ZTKS0213 Z31939 Rev 01 Maintenance ...
Страница 55: ...55 ZTKS0213 Z31939 Rev 01 8 Disassembling Approved persons Bruker Service only ...
Страница 56: ...56 ZTKS0213 Z31939 Rev 01 Disassembling ...
Страница 58: ...58 ZTKS0213 Z31939 Rev 01 Warning Signs ...
Страница 60: ...60 ZTKS0213 Z31939 Rev 01 List of Figures ...
Страница 62: ...62 ZTKS0213 Z31939 Rev 01 List of Tables ...
Страница 64: ...64 ZTKS0213 Z31939 Rev 01 Glossary ...
Страница 80: ...Bruker Corporation info bruker com www bruker com Order No Z31939 ...