121
Z31823_00_05
13 Technical Data
13.1
Voltage Supply Requirements
Input Voltage:
208 -230 VAC (fluctuations of less than + 10%)
Input Frequency Range: 50 - 60 Hz
Pollution Scale:
2
Over Voltage Category: 2
13.2
Site Conditions
The BMPC/2 should only be used:
•
In a standard laboratory environment.
•
Up to a maximum elevation of 2000 meters.
•
At a temperature between 15-35 °C.
•
A relative humidity of a maximum of 80% for temperatures up to 31°C, linearly
decreasing to 67% at a temperature of 35 °C.
13.3
Noise Level
The measured noise level for the BMPC/2 is less than 45 dBA.
13.4
Power Input Safety Fuse
The power input safety fuse for the BMPC/2 is a 50A dual-element time delay fuse Class
J 600V. The Bruker part number for this fuse is
87570
. Two fuses of this type are used in
the unit.
Содержание BMPC/2
Страница 8: ...8 Z31823_00_05 Table of Contents...
Страница 14: ...14 Z31823_00_05 Safety...
Страница 20: ...20 Z31823_00_05 Principles of Operation...
Страница 42: ...42 Z31823_00_05 Alarm System...
Страница 51: ...51 Z31823_00_05 Hardware Overview Figure 8 3 Pump Cabinet V1 V2 P1 P2 Flow meter Oil filter PT100 T5...
Страница 62: ...62 Z31823_00_05 Hardware Overview...
Страница 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Страница 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Страница 122: ...122 Z31823_00_05 Technical Data...
Страница 124: ...124 Z31823_00_05 Contact...
Страница 126: ...126 Z31823_00_05 Warning Signs...
Страница 130: ...130 Z31823_00_05 Figures...
Страница 132: ...132 Z31823_00_05 Tables...
Страница 134: ...134 Z31823_00_05 Index V voice box 43 Voice Box Log 120 Voice files 112 W Wiring Box 54 Z Zoom Function 27...
Страница 135: ......