42
ZTKS0175 / Z31991 / 03
Troubleshooting
6.2.3
During Cooling Down
a. see note on page before
Continued on next page
Indicator
Possible reason
Solution
By
Cooling with liquid
nitrogen continues
too slowly.
Empty transpor-
tation dewar.
Refill or replace transportation
dewar.
Bruker Service
Transfer pressure
too low.
Increase transfer pressure
slightly (max. pressure
0.3 bar).
Bruker Service
Transportation
dewar is leaky; no
transfer pressure
can be applied.
Check transportation dewar
and replace if necessary.
Bruker Service
Precooling with
liquid nitrogen
continues too
quickly.
Transfer pressure
too high.
Stop cooling. Adjust correct
transfer pressure.
Bruker Service
Vacuum in RT
vessel does not
reach
5 x 10
-5
mbar in
48 hours.
O-rings may be
leaky.
Check and clean O-rings and
slots; replace O-rings if
necessary:
•
of the Vacuum Valve
•
of the drop-off plate
•
of the reduction and
sealing flanges
•
of the bottom plate
a
Bruker Service
O-rings may be
frozen due to contact
with liquid nitrogen.
1.
Stop cooling.
2.
Warm up O-ring with
warm air
3.
Wait until the vacuum is
recovered.
4.
Prevent liquid nitrogen
from splashing on
O-rings.
Bruker Service
Defective pumping
unit or pumping line.
Check pumping unit and
pumping line:
A vacuum less than 10
-6
mbar
must be reached with a
closed sealing plug.
Replace if necessary.
Bruker Service
Содержание 600'54 Ascend ULH
Страница 1: ...Version Innovation with Integrity 600 54 Ascend ULH 03 User Manual NMR ...
Страница 4: ...4 ZTKS0175 Z31991 03 ...
Страница 7: ...7 ZTKS0175 Z31991 03 ...
Страница 8: ...8 ZTKS0175 Z31991 03 π ...
Страница 10: ...10 ZTKS0175 Z31991 03 Contact ...
Страница 34: ...34 ZTKS0175 Z31991 03 Transportation ...
Страница 35: ...35 ZTKS0175 Z31991 03 4 Assembling 4 1 Safety Approved Persons Bruker Service only ...
Страница 36: ...36 ZTKS0175 Z31991 03 Assembling ...
Страница 57: ...57 ZTKS0175 Z31991 03 8 Disassembling 8 1 Safety Approved Persons Bruker Service only ...
Страница 58: ...58 ZTKS0175 Z31991 03 Disassembling ...
Страница 59: ...59 ZTKS0175 Z31991 03 A Appendix Warning Signs Figures Tables Glossary Abbreviations Technical Data Index ...
Страница 60: ...60 ZTKS0175 Z31991 03 Appendix ...
Страница 62: ...62 ZTKS0175 Z31991 03 Appendix ...
Страница 64: ...64 ZTKS0175 Z31991 03 Appendix ...
Страница 66: ...66 ZTKS0175 Z31991 03 Appendix ...
Страница 68: ...68 ZTKS0175 Z31991 03 Appendix ...
Страница 74: ...74 ZTKS0175 Z31991 03 Technical Data MS 600 54 Ascend ULH Nitrogen Level Graph Figure A 3 Nitrogen Level Graph ...
Страница 75: ...75 ZTKS0175 Z31991 03 Technical Data MS 600 54 Ascend ULH Helium Level Graph Figure A 4 Helium Level Graph ...
Страница 82: ...82 ZTKS0175 Z31991 03 Technical Data MS 600 54 Ascend ULH This page is intentionally left blank ...
Страница 83: ...83 ZTKS0175 Z31991 03 Technical Data MS 600 54 Ascend ULH Current Lead Figure A 10 Current Lead 55 pins ...
Страница 88: ...88 ZTKS0175 Z31991 03 Technical Data MS 600 54 Ascend ULH This page is intentionally left blank ...
Страница 94: ...94 ZTKS0175 Z31991 03 Technical Data MS 600 54 Ascend ULH This page is intentionally left blank ...
Страница 102: ...Bruker Corporation info bruker com www bruker com Order No ZTKS0175 Z31991 03 ...