54
ZTKS0172 / Z31964 / 04
Troubleshooting
6.3
Troubleshooting Work
6.3.1
After a Quench
Quench while in persistent mode:
1.
Wait until helium stops evaporating and the quench valves are closed.
2.
Wait until there is no helium vapor visible anywhere to make sure there is sufficient
oxygen in the room.
3.
Check that the globes in the quench valves are in the correct position.
4.
Remove probe and shim system to prevent formation of ice on the shim system.
5.
Start the refill with liquid helium as soon as possible.
6.
Contact Bruker Service immediately.
Figure 6.1: Magnet system during a quench
A quench is the very fast de-energizing of
the magnet by loss of its superconductivity.
The stored magnetic energy is converted
into heat, which promotes rapid evaporation
of large quantities of helium.
If a quench occurs contact Bruker Service
immediately.
WARNING
Cryogenic Agents (see
Содержание 500'89 Ascend Aeon
Страница 1: ...Version Innovation with Integrity 500 89 Ascend 04 User Manual NMR ...
Страница 6: ...6 ZTKS0172 Z31964 04 Table of Contents ...
Страница 8: ...8 ZTKS0172 Z31964 04 Contact ...
Страница 28: ...28 ZTKS0172 Z31964 04 Safety ...
Страница 34: ...34 ZTKS0172 Z31964 04 Transportation ...
Страница 35: ...35 ZTKS0172 Z31964 04 4 Assembling Approved Persons Bruker Service only ...
Страница 36: ...36 ZTKS0172 Z31964 04 Assembling ...
Страница 57: ...57 ZTKS0172 Z31964 04 8 Disassembling Approved Persons Bruker Service only ...
Страница 58: ...58 ZTKS0172 Z31964 04 Disassembling ...
Страница 81: ...81 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend 9 21 Current Lead Figure 9 12 Current lead 55 pin ...
Страница 90: ...90 ZTKS0172 Z31964 04 Technical Data MS 500 89 Ascend ...
Страница 92: ...92 ZTKS0172 Z31964 04 Warning Signs ...
Страница 94: ...94 ZTKS0172 Z31964 04 List of Figures ...
Страница 96: ...96 ZTKS0172 Z31964 04 List of Tables ...
Страница 98: ...98 ZTKS0172 Z31964 04 Index ...
Страница 100: ...Bruker Corporation info bruker com www bruker com Order No Z31964 ...