background image

Technical Manual Version 001

BRUKER BIOSPIN

5 (13)

1

Device Information

Introduction

1.1

This document contains technical information about the "BCU-Xtreme External 
Temperature Controller" (

P/N:W1215349

).

The temperature of the gas delivered by the BCU-Xtreme (

P/N:W1212749

) may 

be to too cold for some experiments or some NMR probes, requiring strong 
heating of the cold gas. The new External Temperature Controller Device 
(P/N:W1215349) was designed to gradually adjust the outlet temperature of the 
BCU-Xtreme from -80°C to nearly 0°C. It is compatible with all BCU-Xtreme 
manufactured until 2010.

Safety

1.2

The device is slightly magnetic. The controller must always be handled and 
installed at a safe distance from a magnet.

Controller Installation

1.3

The controller is delivered with several meters of 6mm and 8mm HDPE plastic 
hoses which are connected to the BCU-Xtreme. Optimal controller placement is 
near the BCU-Xtreme or on top of the cooling unit.

Figure 1.1. External temperature controller device top view

Main gas inlet

Temperature
control knob

Содержание BCU-Xtreme

Страница 1: ...NMR Spectroscopy Bruker BioSpin think forward Version BCU Xtreme External Temperature Controller Technical Manual 001...

Страница 2: ...d personnel of the user in charge of repairing the parts of the unit which have been stated by the manufacturer to be repairable Such supply shall in no event constitute permission to modify or repair...

Страница 3: ...IOSPIN 3 13 Contents Contents 3 1 Device Information 5 1 1 Introduction 5 1 2 Safety 5 1 3 Controller Installation 5 1 4 Hoses Connections 6 1 5 Device Operation 6 1 6 Getting Started 7 1 7 Technical...

Страница 4: ...4 13 BRUKER BIOSPIN Technical Manual Version 001 Contents...

Страница 5: ...15349 was designed to gradually adjust the outlet temperature of the BCU Xtreme from 80 C to nearly 0 C It is compatible with all BCU Xtreme manufactured until 2010 Safety 1 2 The device is slightly m...

Страница 6: ...control knob is located on the lid The knob allows you to gradually adjust the outlet temperature from 80 C to approximately to 0 C The knob actuates a couple of valves that divide the inlet gas flow...

Страница 7: ...t is recommended to place the temperature knob in the intermediate position between COLDER and WARMER and then start the operation of the BCU Xtreme to avoid startup problems low gas flow Wait 20 30 m...

Страница 8: ...8 13 BRUKER BIOSPIN Technical Manual Version 001 Device Information...

Страница 9: ...Technical Manual Version 001 BRUKER BIOSPIN 9 13 Figures 1 Device Information 5 Figure 1 1 External temperature controller device top view 5 Figure 1 2 Connections of Controller 6...

Страница 10: ...10 13 BRUKER BIOSPIN Technical Manual Version 001 Figures...

Страница 11: ...Technical Manual Version 001 BRUKER BIOSPIN 11 13 Tables 1 Device Information 5 Table 1 1 Device specification 7...

Страница 12: ...12 13 BRUKER BIOSPIN Technical Manual Version 001 Tables...

Страница 13: ...Technical Manual Version 001 BRUKER BIOSPIN 13 13 End of Document...

Страница 14: ...uker BioSpin Group info bruker biospin com www bruker biospin com Bruker BioSpin your solution partner Bruker BioSpin provides a world class market leading range of analysis solutions for your life an...

Отзывы: