A foundation-based anchorage as well as cross-ply or vibrations damper are not
neccessary.
Cooling aggregate and consumer are generally to be based at the same level.
For other mounting positions- contact the manufacturer.
For fault-free operation of the equipment, as well as the providing corresponding free space
around for maintenance work, the minimum distances recommended are to be adhered to.
In the case of units fitted with rollers, then these rollers are to be secured in position to
protect them from rolling out of position. Many units are actually fitted with 2 rollers with
interlocks to stop them from moving once in position. Press the brake lever downwards to
lock in position.
7.4 General Information
For transportation, we recommend that the equipment’s liquid tank be emptied. For
handling please always keep the equipment in the operating position.
Work on the equipment may only be carried out by trained and qualified persons. The
appropriate safety and environmental protection regulations are to be observed. See
declaration of conformity and/or chapter 1.
The equipment was checked in the works to ensure its tightness.
It has been certified that the item of equipment was subjected to an electrical safety check
in the works before being dispatched. Hence, according to BGV A3, this relieves the
owner/operator of the obligation to carry out a check or to have a check carried out on the
electrical system of the equipment to ensure that it is in the proper working state before
the equipment is put into operation for the first time.
Operating the unit with water temperature below <12°C requires
antifreeze in adequate dosage to avoid damage caused by freezing
up, preferably
Antifrogen N.
Please note that though the insert limit extends downwards the
refrigerating capacity changes considerably. Moreover the pump
capacity has to be checked. Please consult an expert before
reorganizing!