Translation of the Original Operating Instructions
VP-80/VP-80 E
39
6.1.4.2 Charging the battery via the power supply module
To charge the battery fitted in the VIBROPORT 80, use the
AC-7001 power supply module in conjunction with the AC-
1389 USB power supply splitter cable (provided in the scope
of supply).
The AC-1389 splitter cable comprises three connections:
•
USB connector
•
7-pole connector for VP-80 connection USB DEV/TRIG/
PWR
•
Low voltage connector for power supply module
Charge the VP-80:
Connect the two low voltage connectors of the power sup-
ply module and the splitter cable (AC-1389) together.
Connect the splitter cable to the VP-80 (connection USB
DEV/TRIG/PWR) using the 7-pole connector.
Insert the power supply module into a socket.
The VIBROPORT 80 is charged. The red indicator light in-
dicates that the battery is charging.
WARNING!
Explosive atmosphere!
Do not charge the battery of the VIBROPORT 80
E
in potenti-
ally explosive atmospheres. The power supply module and
docking station must not be operated in Ex zones.
This can cause explosions, resulting in damage to property
and injuries.
•
Therefore only charge the battery outside potentially
explosive atmospheres.
•
Never use the power supply module and docking station in
potentially explosive atmospheres.