Chapter 3 – Taking Measurements
Calibration
BB1224 – 11
21
Type 2239 B
Hand-Arm Vibration Functions User Manual
3.1.4
Calibrating Type 2239 B
1.
Make sure the calibration equipment is located away from
all other sources of vibration.
2.
Fit the accelerometer securely onto the calibrator as de-
scribed in your calibrator manual.
3.
Press
to switch on the vibration meter.
4.
Press
to stop the vibration meter from measuring.
5.
Press
on the vibration meter to display the Calibration
Window, as shown below.
6.
Switch on the calibrator. Wait a few seconds for it to warm
up.
7.
Press
to calibrate your instrument. (Press
if you
want to abort the calibration procedure.)
8.
Your instrument is now calibrated.
Note:
If an error message appears after you press
, make sure
that:
●
the calibrator is switched on
●
the accelerometer is correctly and securely attached to
the calibrator
●
the calibrator’s batteries are fully charged
Fig.3.1 The Calibration Window
CALIBRATION
Hand-Arm
Vibration
bb122411.book Page 21 Tuesday, December 11, 2001 2:34 PM