Battery Impedance Meters
BRS
Messtechnik GmbH
Page 17 of 36
Version 1.7 / April2015
If a spectral measurement is started as a sin-
gle scan, the ongoing measurement is indi-
cated by a red bar below. It changes to green
if the measurement cycle is finished.
Frequency grid
The frequency grid can be changed in terms of stop
frequency (1Hz or 0,1Hz) as well as step size (3 values
per decade or 6 values per decade).
Thus speed and resolution can be adapted:
Frequency-Indices
Frequency/Hz 6/D and 0,1Hz
6/D and 1Hz
3/D and 0,1Hz
3/D and 1Hz
1000
1
1
1
1
500
2
2
2
2
333
3
3
250
4
4
3
3
143
5
5
100
6
6
4
4
71,4
7
7
45,5
8
8
5
5
33,3
9
9
21,7
10
10
6
6
15,2
11
11
10,0
12
12
7
7
6,8
13
13
4,7
14
14
8
8
3,3
15
15
2,2
16
16
9
9
1,5
17
17
1
18
18
10
10
0,68
19
0,47
20
11
0,33
21
0,22
22
12
0,15
23
0,10
24
13
Meas. time
ca. 3 min
ca. 22 sec
ca. 2 min
ca. 13 sec
Содержание BIM1