Operational Matrix for the Model DO600
Function /
Resulting Action
Power
Status
Mode
Setting
Required Button Press Sequence
Comments
On/Off
Any
Any
Momentary press of the ON/OFF button
Water Saturated
Air Calibration
On
Any
Place Electrode in Calibration Cap
Press & hold CAL/RECALL button for 2 seconds
Zero Calibration
On Any
Place Electrode in Zero Solution, wait for stability
Press & hold CAL/RECALL button for 2 seconds
Also works with probe
removed
(Cal at zero current)
Store Reading
On
Any
Momentary press of the MODE/HOLD button
Stores and holds
reading “HOLD”
displayed
Hold Release
On
While In
Hold Mode
Momentary press of the MODE/HOLD button
Enter Memory
Retrieval
On Any
Momentary press of the Cal button followed by a
momentary press of the MODE/HOLD button (w/in 4
seconds)
If no data is stored in
memory, “End” is
displayed briefly, and
then returns to last
mode.
Scroll Stored
Readings
On
Memory
Recall
Momentary press of the MODE/HOLD button
Exit Memory
Retrieval
On
Memory
Recall
Momentary press of the CAL/RECALL
Clear Stored
Memory On
Any
Measure-
ment Mode
Press and hold the ON/OFF button for 4 seconds
“clr” is displayed.
Change
Measurement
Mode
On Any
Press and hold the MODE/HOLD button for at least
two seconds (the modes will scroll until the button is
released)
Enter Salinity
Compensation
On Any
Press and release the CAL/RECALL button twice
(momentary) in succession (Displays SAL)
Changing Salinity
Compensation On SAL
Momentary press of the MODE/HOLD button (each
button press increases the ratio by 1 ppt (part per
thousand), the value cycles from 0 to 50)
Exit Salinity
Compensation
On SAL
Press and release CAL/RECALL button for 2 seconds
to enter Altitude Compensation, Or
Press and release Cal Button one more time to enter
measurement mode
It is necessary to
press the
CAL/RECALL button
to save changes. If
the unit times out, no
changes are saved.
Enter Altitude
Compensation
On
Any or SAL
Press CAL/RECALL twice (momentary). Unit enters
Salinity Mode. Press CAL/RECALL for 2 seconds to
enter Altitude Compensation Mode. (Displays Ald)
Unit times out in 5
seconds if no button
is pressed, reverts to
previous mode.
Changing Altitude
Compensation On Ald
Momentary press of the MODE/HOLD button(each
button-press increases altitude by 1,000ft, the value
cycles from 0 to 20)
Exit Altitude
Compensation
On Ald
Momentary press CAL/RECALL button to exit and
save changes.
It is necessary to
press the
CAL/RECALL button
to save changes. If
the unit times out, no
changes are saved.
Change
Temperature
Units
Off
n/a (off
mode)
Press and hold the CAL/RECALL button then
momentarily press the ON/OFF button. Release the
CAL/RECALL button after the "SELF CAL" lights
Override
Autopower Off
On
Any
Press CAL/RECALL button (momentary) then
simultaneously press and hold the MODE/HOLD and
ON/OFF buttons for for 2 seconds.
Default Reset
Off
n/a (off
mode)
Simultaneously press ON/OFF, CAL/RECALL and
MODE/HOLD momentarily. “dFLt” will be displayed.
DO600-EU-EN-V7.0- 8/10
11
Re-order and Accessory Information
Part Number
Description
DO600
ExStik II Dissolved Oxygen Meter
DO605
Replacement Probe, ExStik II Dissolved Oxygen Meter
DO603
Membrane Kit for DO600
Contains: 6 Membrane Caps, 15 mL Filling Solution, Polishing Paper
EX010
Extension Cable 3 feet (1 Meter) & Probe Weight
EX050
Extension Cable 16 feet (5 Meter) & Probe Weight
DO610
ExStik II DO/pH/Conductivity Meter Kit
Contains: EC500 pH/Conductivity/Salinity/TDS ExStik II Meter, DO600
Dissolved Oxygen ExStik II Meter, Single use pH buffer pouches 4, 7,
and 10pH, Sample Cups with Cap, Weighted Base for Sample Cups, and
Batteries, all packed in a Carrying Case
CA895
Small Soft Vinyl Pouch with Belt Loop for ExStik and ExStik II
Copyright © 2010 Extech Instruments Corporation.
All rights reserved including the right of reproduction in whole or in part in any form.