Measurement Units

The meter can be set to measure % saturation, dissolved 

oxygen in mg/l, and dissolved oxygen in parts per million 

(ppm). To change the mode:
1. Press and hold the MODE/HOLD button for 2 seconds 

and the display will begin to scroll through the different 

units of measure : % saturation; D.O. in mg/l; D.O. in 

ppm (parts per million)

2. When the desired units are displayed, release the 

MODE/HOLD button and the unit will return to normal 

operating mode.

Note

: The “HOLD” function cannot be on when changing 

the measurement function. If “HOLD” is displayed in the 

lower left corner of the display, briefly press the MODE/

HOLD button to turn it off.

Temperature Units (ºF/ºC)

1. With the unit OFF, press and hold the CAL/RECALL 

 

button.

2. With the CAL/RECALL button depressed, momentarily 

press the ON/OFF button to turn the unit ON.

3. The CAL/RECALL button can be released when ‘Self Cal’ 

is shown in the display.

Salinity Compensation

1. With the unit ON, momentarily press the CAL/RECALL 

button twice in quick succession (‘SAL’ is shown in the 

lower temperature display).

2. Momentary press the MODE/HOLD button. Each press 

of the MODE/HOLD button increases the compensation 

by 1ppt (part per thousand); the available range is 0 to 

50ppt.

3. Momentarily press the CAL/RECALL button to save the 

compensation setting and return to the normal mea-

surement mode.

Altitude Compensation

1. With the unit ON, momentarily press the CAL/RECALL 

button twice in quick succession (‘SAL’ is shown in the 

lower temperature display.

2. Press and hold CAL/RECALL again for 2 seconds to enter 

the Altitude Mode (‘Ald’ will be shown in the lower tem-

perature display).

3. The factory default value is sea level. Each press of 

the MODE/HOLD button increases the compensation 

by 1000ft. The maximum value is 20 presses (20,000ft 

above sea level).

4. Momentarily press the CAL/RECALL button to save the 

compensation setting and return to the normal mea-

surement mode.

Auto-Power OFF Feature

The auto power off feature automatically shuts the meter 

off 10 minutes after the most recent button press. 

 

To disable this feature, refer to the Disable Auto-Power 

OFF section.

Disable Auto-Power OFF Feature

With the unit ON, momentarily press the CAL/RECALL but-

ton, then quickly press and hold both the MODE/HOLD and 

ON/OFF buttons until ‘off’ is displayed. To restore the Auto 

Power Off Feature (auto power OFF enable) simply turn the 

meter off and on again using the ON/OFF button.

Low Battery Indication

When the battery voltage falls below the operating thres-

hold, ‘BAT’ will appear on the display. Refer to the Mainte-

nance section for battery replacement information.

Storing Readings

1. Press the MODE/HOLD button to store a reading. The 

storage location number will be displayed on the lower 

display, while the main display shows the stored rea-

ding. The meter will enter the HOLD mode and the 

“HOLD” indicator will appear on the LCD.

2. Press the MODE/HOLD button again to exit the HOLD 

mode and return to normal operation. The next time 

MODE/HOLD is momentarily pressed another reading is 

stored and so on.

3. If more than 25 readings are stored, previously stored 

readings (starting with reading number 1) are overwrit-

ten.

Recalling Stored Readings

1. Momentarily press the CAL/RECALL button and then 

with in 4 seconds momentarily press the MODE/HOLD. 

The last stored data point location will be displayed (1 to 

25). Each time the MODE/HOLD button is momentarily 

pressed the next most recently stored data point will be 

displayed.

2. After the last data point stored is displayed, pressing 

the MODE/HOLD button again returns the display to the 

beginning of the list.

3. Pressing the CAL/RECALL button at anytime stops the 

data retrieval process and returns the meter to the nor-

mal measurement mode.

Clearing Stored Readings

1. With the unit ON press and hold the ON/OFF button for 

4 seconds

2. When “clr” is shown in the main display the memory is 

cleared.

Содержание Vinoferm DO-600

Страница 1: ...eling automatische temperatuurcompensatie zoutgehalte en hoogtecompensatie Als u deze meter voorzichtig gebruikt en onderhoudt zal hij u jarenlang trouwe diensten bewijzen Beschrijving van de meter Beschrijving frontpaneel 1 Deksel batterijvak 2 LCD display 3 MODE HOLD toets 4 CAL RECALL toets 5 AAN UIT toets 6 Elektrodekraag 7 Opgeloste zuurstofsensor 8 Borgkapconstructie voor het membraan 9 Memb...

Страница 2: ...was last turned OFF will remain on the display 3 If this is the first time the meter is powered up wait ap prox 3 minutes for the probe to polarize please disable auto power off Instructions for disabling auto power off are provided in a later section For more details please read the Startup Polarization Period paragraph earlier in this manual before continuing 4 Select the desired units of measur...

Страница 3: ...es 20 000ft above sea level 4 Momentarily press the CAL RECALL button to save the compensation setting and return to the normal mea surement mode Auto Power OFF Feature The auto power off feature automatically shuts the meter off 10 minutes after the most recent button press To disable this feature refer to the Disable Auto Power OFF section Disable Auto Power OFF Feature With the unit ON momentar...

Страница 4: ...it disconnects 3 To attach an electrode align the positioning keys on the electrode and the main body housing and then care fully push the electrode into the meter socket until it is fully seated 4 Tighten the electrode retaining collar firmly enough to seal the electrode with the meter DO Bonded Membrane Cap replacement Important Note Do not touch the membrane as skin oils will interfere with the...

Страница 5: ...ne Bonded membrane cap with threaded fitting Operating Temp Range 0 to 50 C 32 to 122 F ATC Range 0 to 50 C 32 to 122 F Salinity Compensation 0 to 50 ppt in 1 ppt part per thousand increments Altitude compensation 0 to 20 000 ft in 1000 ft increments Measurement Storage 25 tagged numbered data sets with recall Battery Power Four 4 CR2032 button batteries Low Battery Indication BAT appears on the L...

Страница 6: ...is depleted Replace electrolyte and membrane cap assembly Readings drift down Insufficient stirring the probe consumes oxygen at the measuring surface requiring constant sample movement across membrane Move probe in sample or stir sample Slow Response Dirty or damaged membrane Replace electrolyte and membrane cap Electrode cannot be calibrated Depleted electrolyte Dirty or damaged membrane Replace...

Страница 7: ...nity Compensation On Any Press and release the CAL RECALL button twice momentary in succession Displays SAL Changing Salinity Compensation On SAL Momentary press of the MODE HOLD button each button press increases the ratio by 1 ppt part per thousand the value cycles from 0 to 50 Exit Salinity Compensation On SAL Press and release CAL RECALL button for 2 seconds to enter Altitude Compensation Or P...

Страница 8: ...ELF CAL lights Override Autopower Off On Any Press CAL RECALL button momentary then simultaneously press and hold the MODE HOLD and ON OFF buttons for for 2 seconds Default Reset Off n a off mode Simultaneously press ON OFF CAL RECALL and MODE HOLD momentarily dFLt will be displayed Re order and Accessory Information Part Number Description DO600 ExStik II Dissolved Oxygen Meter DO605 Replacement ...