1-6
Models SLA5810/20 and SLAMf10/20
Section 1 Introduction
Installation and Operation Manual
X-PR-SLA5800-SLAMf-Series-RevB-PC-eng
Part Number: 541B198AAG
January, 2017
Mark Agency
Certification
Applicable
Standard
Details
UL
(Recogonized)
Class I, Div 2, Group A, B, C, D
Class I, Zone 2, IIC T4
Class II, Zone 22 IP66
UL & CSA Standards
E73889 Vol 3, Sec 4
UL (Listed)
Class I, Div 2, Group A, B, C, D
Class I, Zone 2, IIC T4
Class II, Zone 22 IP66
UL & CSA Standards
E73889 Vol 1, Sec 25
ATEX
II 3 G Ex nA IIC T4 Gc
II 3 D Ex tc IIIC T 85
o
C Dc
IP66
EN
60079
Ͳ
0
:
2012
+
A11
:
2013
EN
60079
Ͳ
15
:
2010
EN
60079
Ͳ
31
:
2014
KEMA 04ATEX1290 X
IECEx
Ex nA IIC T4 Gc
Ex tc IIIC T 85
o
C Dc
IP66
IEC
60079
Ͳ
0
:
2011
+
Corr.
2012
+
Cor.
2013
IEC
60079
Ͳ
15
:
2010
IEC
60079
Ͳ
31
:
2013
IEC KEM 07.0043X
KOSHA
Ex nA IIC T4
Ex tD A22 IP66 T85
o
C
The
Ministry
of
Employment
and
Labor
Notice
No.
2013
Ͳ
34
Article
34
of
the
Industrial
Safety
and
Health
15-AV4BO-0638
15-AV4BO-0639
16-AV4BO-0328X
16-AV4BO-0327X
CE
EMC Directive 2014/30/EU
Directive 2011/65/EU
EN:61326-1:2013 EMC
RoHS
ATEX Special Conditions:
1. When the temperature under rated condition exceeds 70 °C at the cable or conduit entry point, or 80 °C at the
branching point of the conductors, the temperature specification of the selected cable shall be in compliance with
the actual measured temperature values.
2. Provisions shall be made to prevent the rated voltage from being exceeded by transient disturbances of more than
40%.
IECEx Special Conditions:
1. When the temperature under rated condition exceeds 70 °C at the cable or conduit entry point, or 80 °C at the
branching point of the conductors, the temperature specification of the selected cable shall be in compliance with
the actual measured temperature values.
2. Provisions shall be made to prevent the rated voltage from being exceeded by transient disturbances of more than
40%.
3. The equipment shall only be used in an area of not more than pollution degree 2, as defined in IEC 60664-1.
SLAMf UL Listed Installation Instructions:
1. The hazardous locations classes, groups and division as described under products covered.
2. WARNING – EXPLOSION HAZARD – Do not disconnect equipment unless power has been removed or the area
is known to be non-hazardous.
3. WARNING – EXPLOSION HAZARD - Substitution of any components may impair suitability for Class I, Division 2.
4. To maintain type 4X/IP66 rating zero screw must be installed.
UL Recognized versions of SLAMf - Conditions of Acceptability:
1.
These devices are to be used within their ratings.
2.
These devices are to be mounted in a suitable enclosure in the end product rated not less than IP40.
3.
The wiring terminals are suitable for field wiring.
4.
The suitability of the wiring method is to be determined in the end-use application.
5.
These devices are intended to be powered by a class 2 power source.
6.
These devices were evaluated for use in a maximum ambient temperature of 65°C.
KOSHA Special Conditions:
1. The equipment shall only be used in an area of not more than pollution degree 2, as defined in KS C IEC 60664-1.
2. When the temperature under rated condition exceeds 70 °C at the cable or conduit entry point, or 80 °C at the
branching point of the conductors, the temperature specification of the selected cable shall be in compliance with
the actual measured temperature values.
3. The surge protector should be installed to prevent a transient voltage. (140% of rated voltage)
Table 1-5 Certifications - SLAMfXX