![Brooks Polycold MaxCool 2500L Скачать руководство пользователя страница 39](http://html1.mh-extra.com/html/brooks/polycold-maxcool-2500l/polycold-maxcool-2500l_installation-and-operation-manual_2818694039.webp)
Polycold Cryochiller
Safety
Installation and Operation Manual
Mechanical Hazards
Brooks Automation
214072 Revision B
2-13
Lifting Ring Hazard
This unit is equipped with lift rings that when used improperly may
fail and cause death or serious personal injury.
To avoid this hazard, follow these precautions:
•
Lift with 4 straps that each have a minimum strength as required
by local regulations but not less than the weight of the lifted item.
•
Length for each strap between the lift ring bolt and the lifting
point must be greater than 736mm (29 inches).
•
The angle between the sling strap and horizontal at each lift ring
must be greater than 45°
•
Strength of lift device must be greater than 4 times weight of unit
being lifted.
Trip Hazard
Trip hazard which may cause personal injury exists if the facilities
connections such as power and communications cables, gas and
vacuum lines, for the compressor are not properly routed.
•
Route facilities connections and communication cables out of
the way where there is no travel and they do not cause trip haz-
ards
Heavy Object
The Polycold Cryochiller weighs a minimum of 533 kg (1175 lb) up
to a maximum of 544 kg (1200 lb).
Failure to take the proper precautions before moving this unit could
result in personal injury.
•
Follow appropriate methods for moving heavy equipment.