Brooks Automation
122
278580 Revision A
7 Operation
LF80C ZeniD RFID Reader
7.3 Operation of SECS Protocol
Product Manual
7.3.10 SECS-I Message Examples
All examples are produced with the default gateway-ID 255 (decimal) or
0x00FF (hexadecimal)!
S1F1 Message from the reader to the host
Reader to Host: S1F1
10:30:47 Incoming: ENQ ( 05 )
10:30:47 Outgoing: EOT ( 04 )
10:30:47 Incoming: Length Byte ( 0A )
10:30:47 Incoming: Header ( 81 FF 81 01 80 01 00 01 00 01 )
10:30:47 Incoming: Checksum ( 02 85 )
10:30:47 Outgoing: ACK ( 06 )
Host to Reader: S1F2
10:30:47 Outgoing: ENQ ( 05 )
10:30:47 Incoming: EOT ( 04 )
10:30:47 Outgoing: Length Byte ( 10 )
10:30:47 Outgoing: Header ( 01 FF 01 02 80 01 00 01 00 01 )
10:30:47 Outgoing: Data ( 01 02 41 00 41 00 )
10:30:47 Outgoing: Checksum ( 0B 02 )
10:30:47 Incoming: ACK ( 06 )
S1F1 Message from the host to the reader
Host to Reader: S1F1
10:29:00 Outgoing: ENQ ( 05 )
10:29:00 Incoming: EOT ( 04 )
10:29:00 Outgoing: Length Byte ( 0A )
10:29:00 Outgoing: Header ( 01 FF 81 01 80 01 00 00 00 01 )
10:29:00 Outgoing: Checksum ( 04 02 )
10:29:00 Incoming: ACK ( 06 )
Reader to Host: S1F2
10:29:00 Incoming: ENQ ( 05 )
10:29:00 Outgoing: EOT ( 04 )
10:29:00 Incoming: Length Byte ( 1C )
10:29:00 Incoming: Header ( 81 FF 01 02 80 01 00 00 00 01 )
10:29:00 Incoming: Data ( 01 02 41 06 52 53 72 64 30 31 41 06 56 )
10:29:00 Incoming: Data ( 31 2E 30 2E 30 )
10:29:00 Incoming: Checksum ( 05 B5 )
10:29:00 Outgoing: ACK ( 06 )