BRONKHORST® 

 

 

 

 

Page 6 

MASS-VIEW®

 

series 

9.17.051 

 

1.5 

Warranty 

Bronkhorst® products are warranted against defects in material and workmanship for a period of three 
years from the date of shipment, provided they are used in accordance with the ordering specifications 
and the  instructions  in  this  manual  and  that  they  are  not  subjected  to  abuse,  physical  damage  or 
contamination. Products that do not operate properly during this period may be repaired or replaced at 
no  charge.  Repairs  are  normally  warranted  for  one  year  or  the  balance  of  the  original  warranty, 
whichever is the longer. 
The warranty includes all initial and latent defects, random failures, and undeterminable internal causes. 
It  excludes  failures  and  damage  caused  by  the  customer,  such  as  contamination,  improper  electrical 
hook-up, physical shock etc. 
Re-conditioning of products primarily returned for warranty service that is partly or wholly judged non-
warranty may be charged for. 
Mass  Flow  ONLINE  B.V.  prepays  outgoing  freight  charges  when  any  party  of  the  service  is  performed 
under warranty, unless otherwise agreed upon beforehand, however, if the product has been returned 
collect to Bronkhorst High-Tech B.V., these costs are added to the repair invoice. Import and/or export 
charges, foreign shipping methods/carriers are paid for by the customer. 

1.6 

General Description 

The 

MASS-VIEW

®

 

  series  operate  on  the  principle  of  direct  thermal  mass  flow  measurement  (no  by-

pass).  An  advantage  of  using  this  type  of  sensor  is  that  the  instruments  measure  direct  mass  flow, 
virtually independent of pressure and temperature changes, and have a low pressure drop.  
Other advantages compared to conventional  meters are the high accuracy and the large  range ability. 
The 

MASS-VIEW

®

 

 series are supplied with totalizer and alarm functions. Flow control can be achieved 

by  an  optional  built-in  needle  valve.  A  bright  graphical  OLED  display,  clearly  visible  over  any  angle, 
allows reading of actual flow (value and bar graph), total flow and fluid type. The display features easy 
navigation  through  a  user-friendly  menu,  using  a  4-way  button  with  centre  push.  Furthermore  the 
instruments have 2 potential free alarm contacts grounded on one side. 

1.7 

Housing 

The 

MASS-VIEW

®

 

  meter  has  a  straight  tube  with  a  mass  flow  sensor.  The  fluid  connections  have  an 

inner thread of G1/4”  (1/4”  BSPP  cavity) or G1/2”  (G1/2” BSPP  cavity). The connections reside on the 
back of the instrument.  

 

 

The bottom fluid connection is the inlet; the upper connection is the outlet. 

For electrical connection and power supply, the instrument has a RJ-45 modular connector on top of the 
instrument.  These  instruments  are  suited  for  indoor  (dry)  applications,  like  laboratories  and  in  well 
protected (OEM) housings. The front part is made of ABS, the back part of Aluminium. The laminar flow 
element  and  orifice  size  inside  the  control  valve  are  optimized  for  the  customers  gas  and  process 
conditions. 

 
 
 
 
 

MASS-VIEW® 

 

 

MASS-VIEW® 

 

with needle valve 

MASS-VIEW® 

 

with constant-flow controller 

Содержание MASS-VIEW series

Страница 1: ...ries Doc nr 9 17 051G Date 10 04 2018 ATTENTION Please read this instruction manual carefully before installing and operating the instrument Not following the guidelines could result in personal injury and or damage to the equipment ...

Страница 2: ...tion and publication of the contents of this manual we do not assume legal or other liability for any inaccuracy mistake misstatement or any other error of whatsoever nature contained herein The material in this manual is for information purposes only and is subject to change without notice ...

Страница 3: ... gas pressure Electrical connections must be made with a standard cable or according to the hook up diagram in this manual When the instrument is powered up the first time a setup menu appears where the pre installed gas can be selected which flow range is to be used and which communication type Use the 4 way button up and down direction for navigation and press the centre of the button to select ...

Страница 4: ...nd return instructions 10 2 3 Mounting 11 2 4 In line filter 11 2 5 Fluidic connections 11 2 6 Piping 11 2 7 Electrical connections 11 2 8 Test pressure 12 2 9 Supply pressure 12 2 10 System purging 13 2 11 Seals 13 2 12 Equipment storage 13 2 13 Electromagnetic compatibility 13 3 Operation 14 3 1 General 14 3 2 Power and warm up 14 3 3 Setting up the instrument 14 3 4 User Interface menu structur...

Страница 5: ...rmation Discarding this information could cause injuries to people or damage to the instrument or installation Helpful information This information will facilitate the use of this instrument Additional info available on the internet or from your local sales representative 1 4 Product Support References MASS VIEW instruments have modular instruction manuals consisting of Instruction manual MASS VIE...

Страница 6: ...of direct thermal mass flow measurement no by pass An advantage of using this type of sensor is that the instruments measure direct mass flow virtually independent of pressure and temperature changes and have a low pressure drop Other advantages compared to conventional meters are the high accuracy and the large range ability The MASS VIEW series are supplied with totalizer and alarm functions Flo...

Страница 7: ...y of the gas m mass flow 1 9 Needle valve principle The needle valve has an orifice with a long tapered conical seat A needle shaped plunger on the end of a screw exactly fits this seat As the screw is turned and the plunger retracted flow between the seat and the plunger increases Since it takes many turns of the fine threaded screw to retract the plunger precise regulation of the flow rate is po...

Страница 8: ... However when the upstream pressure increases it will cause a small effect on the spring force which means that the ΔP to open the membrane valve will be a little smaller This results in a lower flow In practise the deviation of the ideal straight line will be less than 0 5 1 per bar 1 11 Electronics display and 4 way button The MASS VIEW has a bright graphical OLED display clearly visible over an...

Страница 9: ... DDE provides the user a basic level of inter process communication between MS Windows applications and is therefore the most flexible way to transfer data By using DDE commands sent by software to the DDE Server data can be sent or requested from the FLOW BUS FlowDDE communicates very easily with Microsoft Excel and other applications FlowPlot needs FlowDDE Flow Plot is a DDE client program for s...

Страница 10: ...t www massflow online com and completely fill out the RMA request Instrument handlings Purge gas lines Remove instrument from line Insert the instrument into a plastic bag and seal the bag Place the bag in an appropriate shipping container Add documentation Reason of return Failure symptoms Contaminated condition Declaration on Contamination When returning material always describe the problem and ...

Страница 11: ...ltage and current rating are according to the specifications of the instrument s and furthermore that the source is capable of delivering enough power to the instrument Refer to Hook up MASS VIEW series document nr 9 16 085 for more details When using other cables cable wire diameters should be sufficient to carry the supply current and voltage losses must be kept as low as possible When connectin...

Страница 12: ...n the instrument label The absolute maximum line pressure is 10 bar g Check test pressure before installing in the line If the label is not available or the test pressure is incorrect the instrument should not be mounted in the process line 2 9 Supply pressure Do not apply pressure until electrical connections are made When applying pressure to the system take care to avoid pressure shocks in the ...

Страница 13: ...t of gases at the stated conditions 2 12 Equipment storage The equipment should be stored in its original packing in a cupboard warehouse or similar Care should be taken to not subject the equipment to excessive temperatures or humidity 2 13 Electromagnetic compatibility The MASS VIEW series carries the CE mark Therefore it has to comply with the EMC requirements as are valid for these instruments...

Страница 14: ... minutes for warming up and stabilizing During the warm up period gas pressure may either be on or off 3 3 Setting up the instrument After powering on the first time the instrument will ask which gas is used and which flow range is to be used see picture below Navigate to the fluidname using the 4 way button up and down direction and press in the centre to make a selection then select the range us...

Страница 15: ...ker where the actual flow should be The Flow Marker value can be set in the menu marker 6 Readout parameter 2 is displayed at the bottom of the screen and indicates the actual flow and the selected unit 3 4 2 Changing the readout screen It is possible to customize the readout screen using the navigation button see picture below By pressing left or right one can switch between three different graph...

Страница 16: ...look like the following In the value list edit screen a value can be selected from a list using the up and down button The number of parameter values in the list and the index of the selected item will be displayed on top of the screen In the value edit screen the parameter value can be changed by selecting the digits using left or right and then increase or decrease the value using up or down Pre...

Страница 17: ... Limit Batch limit This parameter is only active when counter mode Up to is selected Reset Resets the counter 3 4 6 Alarms The alarms can be adjusted in the Alarms menu using the following parameters Activate FlowDDE parameter 118 This parameter sets one of the following alarm modes Name DDE Value Description None 0 Alarms Off Min max 1 Alarm on absolute limits Response 2 Alarm on limits related t...

Страница 18: ... are enabled this is made visible in the readout screen with two markers at the left of the graph see picture on the right When an alarm arises the corresponding marker will be highlighted and the header will blink and display the alarm The MASS VIEW meters have two relays which indicate an alarm situation The status of the relays depend on the alarm source and status of the relay parameter see ta...

Страница 19: ...r gases in the heterogeneous gas state where specific heat density thermal conductivity and viscosity can vary tremendously 3 6 Zeroing procedure The MASS VIEW has a zero procedure for each individual gas The zero function permits the user to zero the instrument when needed Please make sure the instrument is zeroed on the desired gas at the desired operating temperature and pressure conditions It ...

Страница 20: ...instrument After cleaning a recalibration is needed Bronkhorst has a trained staff of servicemen available Contact Mass Flow ONLINE B V for cleaning and recalibration options If the equipment is not properly serviced serious personal injury and or damage to the equipment could be the result It is therefore important that servicing is performed by trained and qualified service personnel 4 2 Needle ...

Страница 21: ...ter blocked 1e Clean filter Sensor failure 1f contact online Mass Flow ONLINE Maximum output signal maximum flow indication Flow too high 2a reduce the flow or re range the instrument PCB failure 2b contact online Mass Flow ONLINE Sensor failure 2c contact online Mass Flow ONLINE High setpoint vs output offset flow indication lower than expected Pressure not correct 3a Check pressure Blockage in l...

Отзывы: