background image

04/09/2017 - V1

MEDO220GD & MEDO374GD

Содержание MEDIFRIDGE MEDO220GD

Страница 1: ...04 09 2017 V1 MEDO220GD MEDO374GD...

Страница 2: ...ut prior notification This Installation operation and service manual may not be reproduced in any form with prior written consent from Bromic Refrigeration 1 PREFIX 3 2 IMPORTANT NOTES WARNINGS 4 3 PR...

Страница 3: ...tion manual Any safety information will be accompanied by the following safety alert symbols DANGER WARNING IMPORTANT READ THIS MANUAL CAREFULLY before installing or servicing this product Improper in...

Страница 4: ...h reduced physical sensory or mental capabilities or lack of experience and knowledge unless they have been given supervision or instruction concerning use of the appliance by a person responsible of...

Страница 5: ...wder coated sheet metal Electrical Voltage V 240 240 Frequency Hz 50 50 Input Power W 235 350 Current A 2 2 7 Door Solid or Glass Low E Glass Argon Gas Low E Glass Argon Gas Shape Flat Flat Self Closi...

Страница 6: ...is exposed to warm draughts or open doors Keep the air intake and outlets free of obstructions Adequate allowance should be made for door openings Ventilation Ensure the following instructions are obs...

Страница 7: ...le is rated 70 C It is recommended that the cable should not come into contact with parts over 50 C Do not turn off power before consulting the person responsible Mark the power source clearly so the...

Страница 8: ...rature after the unit has been connected to the power supply The LED indicates that refrigeration circuit is operating During the start up delay 3 minutes the LED is flashing The LED indicates that th...

Страница 9: ...tures two 2 times a day Recordings should be taken before the first door opening of the day MIN Temperature i Press the arrow key ii The Lo message will be displayed followed by the minimum temperatur...

Страница 10: ...advise is given Some products are sensitive to ultraviolet UV light please see storage guidelines from the manufacturer before removing from packaging Do not store food and other goods in the refriger...

Страница 11: ...the products and transfer to a suitable location prior to cleaning the inside of the cabinet Do not use any aggressive or abrasive cleaning agents Never use a water hose or a high pressure jet to cle...

Страница 12: ...ent temperature too high B Inadequate ventilation C Wall clearances below minimum distance D Dirty condenser coil E Ice build up on evaporator coil A Relocate unit B Keep ventilation slots free C Adju...

Страница 13: ...s must be observed In the interest of preventing environmental nuisance and promoting re use of the high quality materials employed used commercial refrigeration equipment must be collected for expert...

Страница 14: ...d storage of any stock Failure to follow instructions may result in loss of or damage to stock Please refer to the National Vaccine Storage Guidelines Strive for 5 for more information on vaccine stor...

Страница 15: ...10 Phiney Place Ingleburn NSW 2565 Australia T AU 1300 276 642 F 61 1300 735 115 E refrigeration bromic com au SYDNEY HEAD OFFICE SHOWROOM BROMIC COM NSW VIC QLD TAS WA SA NT ACT...

Отзывы: