A016 GB Issue 02
13
12.5
The exterior of the incubator may be cleaned with a damp cloth. Avoid allowing any moisture to get
inside electrical housings.
12.6
ALWAYS CLEAN THE INCUBATOR BEFORE STORAGE AND ENSURE THAT THE UNIT IS
TOTALLY DRY INSIDE AND OUT. ALLOW IT TO RUN FOR 24 HOURS WITH NO WATER IN
TO ENSURE IT IS THOROUGHLY DRY.
13 Servicing and Calibration
13.1
In case of failure first check that the mains Power Supply Unit is fully engaged in the socket and that its
lead and plug are fully engaged in the socket in the top of the incubator. The digital control system may
be reset to the original factory defaults by connecting the power supply while holding the OK button.
Check temperature calibration after resetting to defaults.
If the problem persists contact your distributor or Brinsea Products Service Dept.
13.2
The functional parts of the Mini Advance are modular and parts are available and are readily exchanged
by a suitably qualified person equipped with basic tools. Fitting instructions are supplied with
replacement parts.
13.3
The digital temperature display is individually calibrated during manufacture but may be re-calibrated if
required. To ensure optimal performance return the incubator to Brinsea Products Service dept. for re-
calibration every two years. It is not recommended that this procedure is carried out by the user.
BE CAUTIOUS OF LOW COST ANALOGUE OR DIGITAL THERMOMETERS.
BRINSEA PRODUCTS LTD USES SOPHISTICATED EQUIPMENT TRACEABLE TO
INTERNATIONAL REFERENCE STANDARDS.
To access the Calibration Menu press all three buttons simultaneously to unlock the display.