6 Calibration
1.
Turn on the scale and press and hold the
TARE
key during the self-test.
Display will show “Pn”.
2.
Key in the PIN number to get into the scales parameter menu. The default PIN
number is 0000.
3.
Press the
0
numeric key four times…
Pn - - - -
is displayed.
4.
Press the
TARE
key to enter the parameter settings.
menu unit kg
is displayed.
a. If calibrating in kg, continue to step
b. If calibrating in lb, press the
UNITS
key until
lb
is shown on the
display.
5.
Press the
TARE
key to confirm…
f0 cal
is displayed.
6.
Remove all weight from weighing pan and make sure
STABLE
indication is
visible on the display.
7.
Press the
TARE
key to confirm…
unload
is displayed.
8.
Press the
TARE
key to confirm…
CAP loadXX
is displayed.
9.
If necessary, press the
M+
key to select the calibration weight value and then
place the calibration weight on the weighing pan.
10.
When the weight becomes stable, press the
TARE
key to confirm…
PASS
is briefly displayed and then
F0 Cal
.
11.
Calibration is now complete.
12.
Remove calibration weight from the platform.
13.
Press the
ZERO
key to go back to the weighing mode.
Press calibration switch to enable these settings and allow calibration to be
completed.
The calibration switch below the scale may be sealed if legally trade stamped. It
is recommended that recalibration is only carried out by a qualified technician.
Содержание PC3060
Страница 1: ...PC3060 Price Computing Scale User Instructions AWT 35 501856 Issue AA...
Страница 4: ...4 PC3060 Price Computing Scale User Instructions Replacement parts 27 Sealing 28...
Страница 10: ...10 PC3060 Price Computing Scale User Instructions 2 1 Dimensions...
Страница 11: ...PC3060 Price Computing Scale User Instructions 11...
Страница 21: ...PC3060 Price Computing Scale User Instructions 21 3 Press the QTY key to return normal weighing mode...
Страница 29: ......