46
Technical data
Parameter
Value
KVO rate
Q
Rate:
≥
10 ml/h: KVO rate 3 ml/h
Q
Rate: < 10 ml/h: KVO rate 1 ml/h
Q
Rate: < 1 ml/h: KVO rate = rate set using the service
program (factory default rate 0.1 ml/h) or current rate if
this is lower.
History protocol
Q
1,000 history entries
The oldest entries are overwritten if necessary.
Q
100 events for system diagnosis
The history is retained when the device is switched off or
the battery removed.
Delivery rates
Continuous delivery rates/bolus rates according to the syringe size used:
Syringe size [ml]
Continuous
delivery rate [ml/h]
Bolus rate [ml/h]
Preset bolus rate
[ml/h]
50/60
0.01 to 200
Or alternatively:
0.01 to 999.9
1 to 1 800
800
30/35
0.01 to 100
1 to 1 200
600
20
0.01 to 100
1 to 800
400
10/12
0.01 to 50
1 to 500
200
5/6
0.01 to 50
1 to 300
150
2/3
0.01 to 25
1 to 150
80
Note:
The delivery rate can be set in steps of 0.01 ml.
Note:
The preset bolus rate can be changed via the service menu or once via the combination
of bolus volume and bolus time.
Delivery rate accuracy in bolus administration is generally ± 2%. The accuracy can vary
when administering low bolus volumes.