Specifications
Model: BUA5000AR
Operation principle:
Oscillometric method
Display:
Liquid crystal display
Cuff pressure display
range:
0-299 mmHg
Pulse measurement:
40-199 beats per minute
Blood pressure
measurement range:
30-280 mmHg
Laboratory accuracy:
±3 mmHg (cuff pressure)
±5% of reading (pulse rate)
Clinical accuracy:
According to EN 1060-1/-3 and ISO 81060-2
with auscultatory reference:
<5 mmHg systematic offset
<8 mmHg standard deviation
Inflation: Automatic
Power supply:
4 batteries, type AA 1 .5V
Operating temperature/
atmospheric pressure/
humidity:
10 °C to 40 °C (50 ˚F to 104 ˚F)
860 to1060 hPa (0 .849 to 1 .046 atm)
15% to 90% relative humidity, non-condensing
Storage temperature/
atmospheric pressure/
humidity:
-20 °C to 55 °C (-1 ˚F to 131 ˚F)
860 to1060 hPa (0 .849 to 1 .046 atm)
15% to 90% relative humidity, non-condensing
Cuff:
Universal cuff (22-42 cm)
Clinical validation:
According to international protocol 2002 (ESH)
Approximate Service:
5 years
If device is not used within specified temperature and humidity ranges the
technical accuracy cannot be ensured .
Type BF applied parts
Operating temperature
See instructions for use
55
Storage temperature
Keep dry
Subject to change without notice .
Internally powered equipment .
Continuous operation .
English