Rev. 6.4 dated 2022-02
Software Released Version: 1
Part 1 • Page 13-2
Menu selection
Part 1 – Operating Instructions
(Eco)RO Dia II
Part 1 • Chap
ter 13
13.2 Parameters for dialysis mode, menu item 2
The RO device data parameters are set locally during commissioning.
The standard values (factory settings) of the device data to be set are shown here in each sample display
and are adapted locally to the operating conditions and logged during commissioning. If these settings are
changed, it is useful to record the values with the date of change. The entry of new parameter values is
password-protected.
Flow rate:
Volume flows
FISCAL 1 permeate feed in ring piping [l/h]
FISCAL 2 permeate return [l/h]
FISCAL 3 concentrate in channel [l/h]
Yield:
WCF (water conversion factor = water yield of permeate per water
feed) [%]
Consumption [l/h]
Concentrate [l/h]
Ov. quant. of water:
Total volume flow
FISCAL 1 permeate feed [qm]
FISCAL 2 permeate return [qm]
FISCAL 3 concentrate [qm]
Ov. quant. of water Y9:
Concentrate discharge in channel [qm]
Current flow:
Current concentrate discharge in channel via (Y9) [l/h]
Conductivity raw water CIS1 in µS/cm.
IMPORTANT Entering incorrect values can endanger the proper functioning of the control system.
Interval ROII
Press. build-up ROII
Flush time ROII
Interval ROI
Press. build-up ROI
Flush time ROI
Impulse rinse
Limit temperature
Yield
Economy mode RO
Water hardness
In c.o. hard water al
Delay
Overrun dialysis
Temp. discharge
Open Y9 when fault
Cond. alarm by
Maintenance interval
Change prefilter
MB samples
Timer reset
Cell/cond. raw water
<<
ENTER
Selecting the menu item
<< Switches to the previous menu
!Changed data are transferred!
Cursor up/ Change values
Cursor down/ Change values
OK Activate selection
When the key >Prm dialysis mode< is pressed, the key data shown
are listed
The functions and ranges are explained below
Type-specific parameters are only displayed for the respective system
model.
Example:
Interval RO II => only for EcoRO Dia II
Pressure build-up RO I => only for EcoRO Dia I and EcoRO Dia II