Manual cleaning/disinfection

Prior to manual disinfecting, allow water to drip off for a sufficient length of time to prevent dilution of the dis-
infecting solution.

After manual cleaning/disinfection, check visible surfaces visually for residues.

Repeat the cleaning /disinfection process if necessary.

Manual cleaning with immersion disinfection

D–W:

Drinking water

FD–W:

Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)

RT:

Room temperature

*Recommended: BBraun Stabimed

Note the information on appropriate cleaning brushes and disposable syringes, see Validated cleaning and dis-
infection procedure.

Phase I

Fully immerse the product in the cleaning/disinfectant for at least 15 min. Ensure that all accessible surfaces are
moistened.

Clean the product with a suitable cleaning brush in the solution until all discernible residues have been removed
from the surface.

If applicable, brush through non-visible surfaces with an appropriate cleaning brush for at least 1 min.

Mobilize non-rigid components, such as set screws, links, etc. during cleaning.

Thoroughly rinse through these components with the cleaning disinfectant solution (at least five times), using a
disposable syringe.

Phase II

Rinse/flush the product thoroughly (all accessible surfaces) under running water.

Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.

Drain any remaining water fully.

Phase III

Fully immerse the product in the disinfectant solution.

Mobilize non-rigid components, such as set screws, joints, etc. during rinsing.

Rinse lumens at least 5 times at the beginning of the exposure time using an appropriate disposable syringe.
Ensure that all accessible surfaces are moistened.

Phase IV

Rinse/flush the product thoroughly (all accessible surfaces).

Mobilize non-rigid components, such as set screws, joints, etc. during final rinse.

Rinse lumens with an appropriate disposable syringe at least five times.

Drain any remaining water fully.

Phase V

Dry the product in the drying phase with suitable equipment (e.g. cloth, compressed air), see Validated cleaning
and disinfection procedure.

Mechanical cleaning/disinfecting

Note

The cleaning and disinfection device must be of tested and approved effectiveness (e.g. FDA approval or CE mark
according to DIN EN ISO 15883).

Note

The cleaning and disinfection device used for processing must be serviced and checked at regular intervals.

Mechanical alkaline cleaning and thermal disinfecting

Machine type: single-chamber cleaning/disinfection device without ultrasound

D–W:

Drinking water

FD–W:

Fully desalinated water (demineralized, low microbiological contamination: drinking water quality
at least)

*Recommended: BBraun Helimatic Cleaner alkaline

Check visible surfaces for residues after mechanical cleaning/disinfecting.

Inspection, maintenance and checks

Allow the product to cool down to room temperature.

After each complete cleaning, disinfecting and drying cycle, check that the product is dry, clean, operational, and
free of damage (e.g. broken insulation or corroded, loose, bent, broken, cracked, worn, or fractured components).

Dry the product if it is wet or damp.

Repeat cleaning and disinfection of products that still show impurities or contamination.

Check that the product functions correctly.

Immediately put aside damaged or inoperative products and send them to Aesculap Technical Service,  see Tech-
nical Service.

Check for compatibility with associated products.

Packaging

Place the product in its holder or on a suitable tray. Ensure that all cutting edges are protected.

Pack trays appropriately for the intended sterilization process (e.g. in Aesculap sterile containers).

Ensure that the packaging provides sufficient protection against recontamination of the product during storage.

Steam sterilization

Note

The product can be sterilized either in disassembled or in assembled condition.

Check to ensure that the sterilizing agent will come into contact with all external and internal surfaces (e.g. by
opening any valves and faucets).

Validated sterilization process

Steam sterilization using fractional vacuum process

Steam sterilizer according to DIN EN 285 and validated according to DIN EN ISO 17665

Sterilization using fractional vacuum process at 134 °C/holding time 5 min

When sterilizing several products at the same time in a steam sterilizer, ensure that the maximum load capacity
of the steam sterilizer specified by the manufacturer is not exceeded.

Sterilization for the US market

Aesculap advises against sterilizing the device by flash sterilization or chemical sterilization.

Sterilization may be accomplished by a standard prevacuum cycle in a steam autoclave.

To achieve a sterility assurance level of 10

-6

, Aesculap recommends the following parameters:

*Aesculap has validated the above sterilization cycle and has the data on file. The validation was accomplished in an
Aesculap sterile container cleared by FDA for the sterilization and storage of these products. Other sterilization cycles
may also be suitable, however individuals or hospitals not using the recommended method are advised to validate
any alternative method using appropriate laboratory techniques. Use an FDA cleared accessory to maintain sterility
after processing, such as a wrap, pouch, etc.

Storage

Store sterile products in germ-proof packaging, protected from dust, in a dry, dark, temperature-controlled area.

Technical Service

For service and repairs, please contact your national B. Braun/Aesculap agency.

Modifications carried out on medical technical equipment may result in loss of guarantee/warranty rights and for-
feiture of applicable licenses.

Service addresses

Aesculap Technischer Service
Am Aesculap-Platz 
78532 Tuttlingen / Germany
Phone:

+49 (7461) 95-1602

Fax:

+49 (7461) 16-5621

E-Mail:

[email protected]

Or in the US: 
Aesculap Implant Systems LLC 
Attn. Aesculap Technical Services 
615 Lambert Pointe Drive 
Hazelwood 
MO, 63042 
Aesculap Repair Hotline 
Phone:

+1 (800) 214-3392 

Fax:

+1 (314) 895-4420

Other service addresses can be obtained from the address indicated above.

Disposal

Adhere to national regulations when disposing of or recycling the product, its components and its packaging!

Distributor in the US/Contact in Canada for product information and 
complaints

Aesculap Implant Systems LLC 
3773 Corporate Parkway 
Center Valley, PA, 18034,
USA

TA-Nr. 012480     01/13     V6     Änd.-Nr. 45746

Phase

Step

T
[°C/°F]

t
[min]

Conc. 
[%]

Water 
quality

Chemical

I

Disinfecting 
cleaning

RT (cold)

>15

2

D–W

Aldehyde-free, phenol-free, 
and QUAT-free concentrate, 
pH ~ 9*

II

Intermediate 
rinse

RT (cold)

1

-

D–W

-

III

Disinfection

RT (cold)

15

2

D–W

Aldehyde-free, phenol-free, 
and QUAT-free concentrate, 
pH ~ 9*

IV

Final rinse

RT (cold)

1

-

FD-W

-

V

Drying

RT

-

-

-

-

Phase

Step

T
[°C/°F]

t
[min]

Water 
quality

Chemical/Note

I

Prerinse

<25/77

3

D–W

-

II

Cleaning

55/131

10

FD-W

Concentrate, alkaline: 

pH  =  13

<5 % anionic surfactant

0.5 % working solution

pH = 11*

III

Intermediate 
rinse

>10/50

1

FD-W

-

IV

Thermal disin-
fecting

90/194

5

FD-W

-

V

Drying

-

-

-

According to the program for cleaning 
and disinfection device

CAUTION

Damage (metal seizure/friction corrosion) to the product caused by insufficient 
lubrication!

Prior to function checks, lubricate moving parts (e.g. joints, pusher compo-
nents and threaded rods) with maintenance oil suitable for the respective ster-
ilization process (e.g. for steam sterilization: Aesculap STERILIT® I oil spray 
JG600 or STERILIT® I drip lubricator JG598).

Aesculap Orga Tray/Sterile container (perforated bottom)
Minimum cycle parameters*

Sterilization method

Temp.

Time

Minimum drying time

Prevacuum

270 °F/275 °F

4 min

20 min

WARNING

Risk of injury and/or malfunction!

Do not modify the product.

Содержание Aesculap Metha

Страница 1: ...para cone 12 14 ND656R Gebruiksaanwijzing Technische beschrijving Metha uitslaginstrument voor conus 12 14 ND656R Bruksanvisning Teknisk beskrivning Metha utslagningsinstrument f r kon 12 14 ND656R M...

Страница 2: ...this The recommended chemistry was used for validation Note If there is no final sterilization then a virucidal disinfectant must be used Note For the latest information on reprocessing and material...

Страница 3: ...t protection against recontamination of the product during storage Steam sterilization Note The product can be sterilized either in disassembled or in assembled condition Check to ensure that the ster...

Страница 4: ...erfolgt muss ein viruzides Desinfektionsmittel verwendet werden Hinweis Aktuelle Informationen zur Aufbereitung und zur Materialvertr glichkeit siehe auch Aesculap Extranet unter www extranet bbraun c...

Страница 5: ...en auf Trockenheit Sauberkeit Funktion und Besch digung z B Isolation korrodierte lose verbogene zerbrochene rissige abgenutzte und abgebrochene Teile Nasses oder feuchtes Produkt trocknen Unsauberes...

Страница 6: ...rque Si aucune st rilisation finale n a lieu des produits de d contamination virocides doivent tre utilis s Remarque Pour des informations actuelles sur le traitement st rile et la compatibilit avec l...

Страница 7: ...er sur le produit les l ments suivants tat sec pro pret bon fonctionnement et absence de d t riorations p ex d t rioration de l isolation ou pi ces corrod es l ches tordues bris es fissur es us es et...

Страница 8: ...las sustancias qu micas recomendadas Nota Deber utilizarse un agente de limpieza virucida si no va a esterilizarse el producto a continuaci n Nota Para consultar informaci n actualizada sobre c mo tra...

Страница 9: ...con corrosi n sueltas deformadas rotas agrietadas desgastadas o fragmentadas Secar el producto si est h medo o mojado Si el producto est sucio volver a limpiarlo y desinfectarlo Comprobar el funciona...

Страница 10: ...a responsabilit di ci ricade sul gestore preparatore Per la validazione stata utilizzata la chimica raccomandata Nota Se non vi alcuna sterilizzazione successiva occorre utilizzare un disinfettante vi...

Страница 11: ...d es all isolamento nonch componenti corrosi staccati deformati rotti crepati usurati o altrimenti alterati Asciugare il prodotto bagnato o umido Pulire e disinfettare nuovamente il prodotto non pulit...

Страница 12: ...Nota Para informa es actuais sobre o reprocessamento e a compatibilidade dos materiais ver tamb m a Extranet da Aesculap em www extranet bbraun com O m todo homologado de esteriliza o a vapor foi efe...

Страница 13: ...produto se estiver molhado ou h mido Limpar e desinfectar novamente o produto caso apresente sujidade Verificar se o produto funciona correctamente Eliminar imediatamente os produtos que apresentem da...

Страница 14: ...alidering werden de aanbevolen chemische middelen gebruikt Opmerking Indien geen afsluitende sterilisatie plaatsvindt moet een viricide desinfectiemiddel worden gebruikt Opmerking Actuele informatie o...

Страница 15: ...tie corrosie losse verbogen gebroken gebarsten versleten en afgebroken onderdelen Laat natte of vochtige producten drogen Reinig en desinfecteer verontreinigde producten nogmaals Controleer de werking...

Страница 16: ...de kemikalien Tips Om ingen avslutande sterilisering genomf rs m ste ett desinfektionsmedel med virucid verkan anv ndas Tips Aktuell information om beredning och materialkompatibilitet finns p Aescula...

Страница 17: ...l sa b jda trasiga spruckna utslitna och avbrutna delar Torka v ta eller fuktiga produkter Reng r produkter som inte r rena p nytt och desinficera dem Kontrollera att produkterna fungerar Sortera gena...

Страница 18: ...Metha Metha ND655R Metha 12 14 ND656R ND655R 6 1 close Metha NF275R Metha Metha Metha 1 open 5 1 4 Metha 1 open Metha ND655R Metha ND655R 3 ND656R 1 open 5 1 4 Metha 1 open Metha Metha 1 close 2 5 1...

Страница 19: ...elimatic Cleaner alcaline Aesculap Aesculap DIN EN 285 DIN EN ISO 17665 134 C 5 20 T C F t I 15 2 pH 9 II 1 III 15 2 pH 9 IV 1 V T C F t I 25 77 3 II 55 131 10 pH 13 5 P 0 5 pH 11 III 10 50 1 IV 90 19...

Страница 20: ...B Braun Aesculap Aesculap Technischer Service Am Aesculap Platz 78532 Tuttlingen Germany Phone 49 7461 95 1602 Fax 49 7461 16 5621 E Mail ats aesculap de TA Nr 012480 01 13 V6 nd Nr 45746...

Страница 21: ...pravu K validov n byly pou ity doporu en chemik lie Upozorn n Pokud se neuskute n z v re n sterilizace je nutno pou t virucidn desinfek n prost edek Upozorn n Aktu ln informace k prav a materi lovou s...

Страница 22: ...izolace zkorodovan voln ohnut rozbit praskl opot eben a ulomen sti Mokr nebo vlhk v robek vysu te Zne i t n v robek znovu vy ist te a desinfikujte Zkontrolujte fungov n v robku Po kozen anebo nefunk...

Страница 23: ...et bbraun com Sterylizacj w oparciu o walidowan metod przeprowadzono w systemie pojemnik w sterylnych Aesculap Og lne wskaz wki Zaschni te lub przylegaj ce do urz dzenia pozosta o ci pooperacyjne mog...

Страница 24: ...np izolacji skorodowanych lu nych wygi tych z amanych poryso wanych zu ytych lub od amanych cz ci Mokry lub wilgotny produkt nale y osuszy Zabrudzony produkt ponownie wyczy ci i zdezynfekowa Sprawdzi...

Страница 25: ...iu sa pou va doporu en ch mia Ozn menie Ak nenasleduje na z ver steriliz cia mus by pou it virucidn dezinfek n prostriedok Ozn menie Pre aktu lne inform cie o pr prave a kompatibilite materi lu pozri...

Страница 26: ...uchos istotu funk nos a po kodenie napr izol ciu skorodovan uvo nen ohnut rozbit opotrebovan a odlomen kusy Vlhk alebo mokr v robok vysu i Zne isten v robok znova vy isti a dezinfikova Skontrolova fun...

Страница 27: ...ekle medi inde bir vir sidal dezenfeksiyon maddesi kullan lmal d r Not Haz rl k ve malzeme uyumlulu una y nelik g ncel bilgiler i in bak n zAesculap Extranet www extranet bbraun com Do rulanan buharla...

Страница 28: ...n oda s cakl na so umas n bekleyiniz Her temizlik dezenfeksiyon ve kurutmadan sonra r nde a a daki hususlar kontrol edin Kuruluk temizlik al ma ve izolasyon paslanma gev eme e ilme par alanma y rt lm...

Отзывы: