7
6.
Working with the motor hose with
hand control and with the wall adapter
6.1
Function check
►
Check the ELAN 4 air motor system with a trial run prior to each appli-
cation.
►
Make certain that the ELAN 4 air applied part and ELAN 4 air motor
hose with hand control are in good working order.
►
Make certain that the ELAN 4 air motor hose with hand control and
the wall adapter are correctly connected and leak-tight.
►
Check the secure connection of all products used.
►
Make certain that the applied part runs at maximum speed when the
hand lever of the ELAN 4 air motor hose with hand control is fully acti-
vated.
6.2
System set-up
Note
Operating elements on the system components of the ELAN 4 air motor
system are labeled with a gold marking.
Connecting the accessories
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
►
Follow the instructions for use of individual accessories.
►
Please contact your B. Braun/Aesculap Partner or Aesculap Technical
Service, address see Technical Service.
Connecting wall adapter to compressed-air source
►
Couple connection for compressed-air source
8
of the wall adapter
with the compressed-air source.
Connecting the motor hose with hand control to the wall adapter
►
Attach the connection for wall adapter
1
of the motor hose with hand
control to the connection for motor hose
9
of the wall adapter and
couple the bayonet lock.
Connecting the applied part to the hand control
►
Attach the connection for motor hose of the applied part
11
to the
connection for applied part
5
of the motor hose, see Fig. A. When
doing so, ensure that the nib
6
on the motor hose is aligned with the
groove on the connection for motor hose
10
.
The applied part engages. When the visual field is “Off”
3
on the motor
coupling, a gold-colored marking is visible.
Note
The applied part attached to this motor hose with hand control is only
ready for operation when the release button is in "On" position.
Releasing the applied part for operation ("On" position)
►
Move the hand lever
7
180°.
►
Activate the release button
4
on the motor hose and push the connec-
tion for applied part
5
onto the applied part, see Fig. A.
The applied part engages. When the visual field is “On”
2
on the motor
coupling, a gold-colored marking is visible.
►
Return the hand lever
7
to its initial position.
Function checks
►
To check the motor hose is correctly fitted: Pull the motor hose away
from the applied part at the connection for applied part
5
. Do not acti-
vate the release button
4
.
The motor hose must be securely fitted to the applied part and must
not come loose.
Locking the applied part ("Off" position)
►
Move the hand lever
7
180°.
►
Activate the release button
4
on the motor hose and detach the motor
hose from the applied part.
The applied part engages. When the visual field is “Off”
3
on the motor
coupling, a gold-colored marking is visible.
Disconnecting the applied part from the motor hose with hand control
►
Activate the release button
4
on the motor hose and pull at the con-
nection for applied part
5
to disconnect the motor hose from the
applied part.
WARNING
Risk of infection and contamination!
Applied parts, motor hose and sometimes also tools
are delivered unsterile!
►
Carry out sterile processing of the applied parts,
motor hose and unsterile tools prior to use, in
accordance with the instructions for use.
WARNING
Damage to the product if dropped!
►
Use the products only if they are in perfect
technical condition, see Function check.