9
Only process chemicals that have been tested and approved (e.g. VAH or
FDA approval or CE mark) and which are compatible with the product’s
materials according to the chemical manufacturers’ recommendations
may be used for processing the product. All the chemical manufacturer's
application specifications must be strictly observed. Failure to do so can
result in the following problems:
■
Material damage such as corrosion, cracks, fracturing, premature
aging, or swelling.
►
Do not use metal cleaning brushes or other abrasives that would dam-
age the product surface and could cause corrosion
►
For further detailed information on hygienically safe and material-pre-
serving/value-preserving reprocessing, see www.a-k-i.org, link to Pub-
lications, Red Brochure – Proper maintenance of instruments.
7.3
Preparations at the place of use
►
Separate the products from each other immediately after use.
►
Remove any visible surgical residues as much as possible with a damp,
lint-free cloth.
►
Place the dry product in a sealed waste container and forward it on for
cleaning and disinfection within 6 hours.
►
ELAN 4 air wall adapter: Perform non-fixating/NaCl-free pre-cleaning
immediately after use.
7.4
Preparation before cleaning
ELAN 4 air motor hose with hand control
►
Prior to first mechanical cleaning/disinfection process: Assemble
ECCOS holder GB053R in a suitable tray (e.g. JF222R).
►
Place the product in the correct position in the ECCOS holder GB053R,
see Fig. B and Fig. C.
7.5
Cleaning/disinfection
Product-specific safety instructions for the reprocessing procedure
ELAN 4 air motor hose with hand control
Note
The indicated drying time is a guide time only. It must be checked taking
into account the specific conditions (e.g. load) and if applicable adjusted.
ELAN 4 air wall adapter
CAUTION
Damage to the product due to inappropriate clean-
ing/disinfecting agents and/or excessive tempera-
tures!
►
Following the manufacturer's instructions, use
cleaning and disinfecting agents which
–
are approved for plastic material and high-
grade steel,
–
do not attack softeners (e.g., in silicone).
►
Do not use cleaning agents that contain ace-
tone.
►
Observe specifications regarding concentration,
temperature and exposure time.
►
Do not exceed the maximum temperature of
60 °C during chemical cleaning and/or disinfec-
tion.
►
Do not exceed the maximum temperature of
90 °C during thermal disinfection with FD
water.
►
Dry the product for at least 10 minutes at a
maximum of 120 °C.
CAUTION
Damage to, or destruction of the product caused by
mechanical cleaning/disinfection!
►
Only clean and disinfect the product manually.
►
Do not sterilize the product under any circum-
stances.
CAUTION
Damage to the product due to inappropriate clean-
ing/disinfecting agents!
►
Only use cleaning/disinfecting agents approved
for surface cleaning. Follow the manufacturer’s
instructions for the respective cleaning/disin-
fecting agent.