Aesculap Power Systems
Acculan® 3Ti Dermatome GA670
12
4.2
Function checks
Note
The function checks must be carried out prior to each
surgical application of the unit and after each
intraoperative battery change!
¾
Follow the advisories/procedure for inspection,
maintenance and checks, see Inspection,
maintenance and checks.
¾
Ensure that a dermatome blade is correctly seated
in the dermatome.
¾
Ensure that there is a battery in the battery
compartment.
¾
Check that the roll pin is securely seated.
¾
Check that the dermatome blade is securely seated.
¾
Briefly activate the dermatome at maximum
oscillation frequency.
4.3
Safe operation
Adjusting the cutting depth of the Acculan®
3Ti Dermatome
Note
The skin condition (e.g. due to age) of the patient must
be considered when adjusting the cutting depth (graft
thickness)!
¾
Set the cutting depth at setting lever
14
.
Lock
15
of setting lever
14
prevents inadvertent
changes of the cutting depth setting.
¾
Read the cutting depth (graft thickness) on dial
wheel
16
(1/10 mm gradation).
¾
Follow the troubleshooting advisories/procedure,
see Troubleshooting list.
Adjusting the cutting width of the Acculan®
3Ti Dermatome
Note
The width of the skin graft will be approx. 2 mm less
than the cutting width set by the flaps.
WARNING
Risk of injury caused by inadvertent
activation of the motor!
¾
Secure dermatomes that are not
actually used against
inadvertent activation.
WARNING
Risk of injury and/or malfunction!
¾
Always carry out a function
check before using the product.
WARNING
Risk of injury/cuts by the
dermatome blade if the flaps are
opened incorrectly!
¾
Always use the flap tabs for
lifting the flaps.
Содержание Acculan 3Ti
Страница 2: ...23 26 25 21 22 22 20 24...
Страница 3: ......
Страница 209: ...207 2 7 6 24 10 9 3 1 20 3 24 9 7 6 20 1...
Страница 210: ...Aesculap Power Systems Acculan 3Ti GA670 208 4 23 20 26 5 20 5 23 20 26 20...
Страница 211: ...209 6 21 22 6 7 21 22...
Страница 212: ...Aesculap Power Systems Acculan 3Ti GA670 210 8 9 1 22 21 21 7 23 8...
Страница 213: ...211 9 1 21 25 9 1 25 23 25 25 23 23 10 1 4 4 5 OFF 10 4 1 4 5 ON 10 4 1 25 1 4 5 1...
Страница 214: ...Aesculap Power Systems Acculan 3Ti GA670 212 11 Eccos Eccos 5 Eccos 1 Eccos 4 2 4 3 BH MAH BH MAH...
Страница 215: ...213 Acculan 3Ti 14 15 14 16 1 10 Acculan 3Ti 2 12 12 12 11 12 8 Acculan 3Ti Acculan 3Ti 4 BH MAH 12 11...
Страница 216: ...Aesculap Power Systems Acculan 3Ti GA670 214 Acculan 3Ti 1 1 13 1 5 Aesculap Extranet www aesculap extra net...
Страница 217: ...215 5 1 5 2 6 45 C VAH DGHM FDA CE pH 8 BH MAH GB228R...
Страница 218: ...Aesculap Power Systems Acculan 3Ti GA670 216 www a k i org 5 3 Acculan 3Ti Acculan 3Ti 5 4 30 5 5 55 C...
Страница 219: ...217 5 6 I II III T C F t I II III 1 Meliseptol HBV 50 1 IV 0 5 V...
Страница 223: ...221 6 14 15 2012 6 B Braun Aesculap 14 Acculan 3Ti 15...
Страница 227: ...225 11 B Braun Aesculap Extranet PDF...