temperature probe T1 (standard) and point B for probe T2 should your Tankmaster be 
configured for a two probe system. 
 
To calibrate the controller insert a small flat ended screwdriver into the correct hole until 
it engages with the potentiometer slot below. You can now rotate the screwdriver CW and 
CCW to increase or decrease the displayed temperature. It is recommended you use a 
glass  thermometer  as  a  reference  temperature  to  perform  this  operation  as  detailed 
below. Also ensure that you do not apply undue pressure once you reach the end stops 
on the potentiometers otherwise damage will occur 

 

5.7  Immersion Method Against Known Temperature. 

 

Place the probe and a thermometer in a container of ice and water and allow to settle for 
approx. 5 minutes. 
 
Adjust  potentiometer  A  or  B  until  the  Tankmaster  display  reads  the  same  as  the 
thermometer. 
 
Place the thermometer and the sensor in a second container with water at an elevated 
temperature of say 70

0

C and allow to settle (5 minutes) 

 
Adjust  potentiometers  A  or  B  until  the  Tankmaster  display  reads  the  same  as  the 
thermometer. 
 
Repeat  the  above  procedure  until  satisfied  the  display  is  stable.    The  unit  is  now 
calibrated.  
 
NB The above procedures are recommendations and do not constitute part of a certifiable 
test.    It  is  recommended  that  you  contact  a  standard  testing  centre  if  certification  is 
required.  

 

 

 

 
 
 

Содержание TANKMASTER MP

Страница 1: ...E Combined Temperature and Liquid Level Controller INSTALLATION OPERATION AND MAINTENANCE INSTRUCTIONS These instructions should be carefully read and understood by all operators before installing and...

Страница 2: ...roller capable of controlling liquid and temperature parameters within a process Fig 1 Tankmaster controller Wall mounting holes 4 corners Calibration Pots A B Control LED s Levelmaster HPD probe plug...

Страница 3: ...olt free relays UNLESS specified at the time of order Should the Tankmaster MP be ordered as a mains output voltage type the grey output lead marked with a red warning label and fitted with a protecti...

Страница 4: ...on Black 3 Alarm relay common This relay is volt free and any voltage between 12 and 240v may be applied across it Black 4 Alarm relay output Black 5 Level control output pump This is a volt free rela...

Страница 5: ...n in Fig 2 on page 6 has two LED displays six LED s and three push buttons The larger of the two displays referred to as the main display throughout this document displays the temperature of probe T1...

Страница 6: ...Setting the SET POINT a Press P1 once to display the SET POINT which is shown in the main LED display as SP The actual set point value is shown in the small LED display window below it b Press P2 to...

Страница 7: ...he end stops on the potentiometers otherwise damage will occur 5 7 Immersion Method Against Known Temperature Place the probe and a thermometer in a container of ice and water and allow to settle for...

Страница 8: ...ALISTS Fig 2 Temp and Liquid level status TEMPERATURE LEVEL HEATER PUMP Alarm Low Low Alarm Temperature below set point OFF ON ON Temperature above set point OFF OFF OFF Liquid level low ON OFF OFF ON...

Страница 9: ...Page 9 Date 06 09 2019 IN037 Rev 6 May 2019 THERMAL PROCESS EQUIPMENT SPECIALISTS 6 Wiring connection Fig 3 Wiring diagram for Tankmaster connected to Polaris heater Live relay...

Страница 10: ...the solution level The probe is supplied with a two or three pin IP66 standard plug which should be connected to the appropriate sockets on the underside of the controller Ensure that the locking rin...

Страница 11: ...velmaster probe 10 1 1 Low level only Cut the red sensor to the low level required Cut the white sensor 50mm higher than the low level sensor Cut the green sensor to not less than 30mm below the lowes...

Страница 12: ...s supplied with your immersion heater If the HPD is used to control more than one heater the POLARIS fitted with an HPD probe must be mounted slightly higher than the others Plug the probe lead into t...

Страница 13: ...ng add a small amount of common salt to increase the conductivity of the solution It is recommended that the probe assemblies are cleaned as part of the regular maintenance schedule to avoid build up...

Отзывы: