Assemble and connect the new fan and the cable with electronic circuit
Take the pre-assembled fan with metal base with hoses fittings
Using the plastic rivets (previously disassembled) assemble the fan
onto the case of equipment
Logic board
“pwr prn”
fan
printer
Logic board
“exhaust oil”
ATTENTION –VERIFY THAT THE AIR FLOW DIRECTION IS
DIRECTED TO THE OUTSIDE OF EQUIPMENT CASE
AIR FLOW DIRECTION
(from inside to outside)