           ER900 and ER900 AF Event Monitors 

13

 

 

Troubleshooting 

 

Symptom 

Recommended solution

 

Install batteries or check battery direction. 
Ensure patient electrodes/leads are connected to 
patient properly. 
Ensure patient cable is inserted completely. 
Use only a one channel cable. 

No single beep when inserting 
patient cable 

Lead loss is in off position, call technician. 
Memory full-Phone Ring.  Follow instructions To 
Send and Erase Events. 
Ensure patient cable is inserted completely. 

Will not record 

Ensure RECORD/SEND button is held for two 
seconds. 

Siren (alternating) tone while 
recording 

There is not a good connection. Check that 
electrodes/leads have a good connection to patient 
and cable is plugged into monitor 

Phone ring sound every minute for 
10 minutes 

Memory is full, follow instructions To Send and 
Erase Events. 

Phone ring sound when 
RECORD/SEND button is pushed 

Memory is full, follow instructions To Send and 
Erase Events. 

Phone ring sound once  

An event is already stored in memory at start up, 
also heard at the end of a recorded event.  Follow 
instructions To Send and Erase Events. 

Three beeps every five minutes 
with cable inserted 

Batteries are low.  Replace batteries and/or clean 
battery contacts 
Make sure

 mouthpiece

 of phone is directly over 

monitor speaker 

No information received by 
receiving center 

Ensure RECORD/SEND button is held for two 
seconds. 
Mouthpiece of phone must be close to the monitor 
speaker hole. 
Check telephone connection.  Listen to phone line 
before sending event(s) to ensure there is no noise. 
Have patient call back and send ECG again. 

Noise artifact on recorded ECG at 
receiving center 

Have patient try another phone. 
Electrodes must be securely attached to patient. 
Patient should remain still while recording. 
Replace patient cable. Pulling on lead wires may 
damage cable. 
Verify the recording did not take place near a source 
of electromagnetic interference (fluorescent lights, 
computer monitors, or household appliances). 

Noise artifact on recorded ECG at 
patient location 

Move electrodes slightly to the right or left of the 
original location. 

All or groups of timestamps for 
recordings are the same. 

If the inadvertent loss of power occurs, all the 
timestamps in the FSK will reset to the time the unit 
powered back up.  Subsequent recordings will have 
time stamps relative to the power up time. 

Falling tone 

Transmission complete 

Rising tone 

Ready to record 

 

 

 

           ER900 and ER900 AF Event Monitors 

14

 

Event Markings 

 
The following shows typical event markings that are generated by the 
monitor and appear on the receiving station strip chart. 
 

 

Содержание ER900

Страница 1: ...Braemar Inc ER900 ER900 ER900 ER900 and and and and ER900 AF ER900 AF ER900 AF ER900 AF Arrhythmia Event Monitor Caution U S Federal law restricts this device to sale by or on the order of a physician ...

Страница 2: ...t for the express warranties stated above Braemar disclaims all warranties including implied warranties of merchantability and fitness The stated express warranties are in lieu of all obligations of liabilities on the part of Braemar for damages including but not limited to special indirect or consequential arising out of or in connection with the use or performance of Braemar products Any action ...

Страница 3: ...trode Application and Patient Hookup 9 Patient Operating Instructions 10 This page intentionally left blank 12 Troubleshooting 13 Event Markings 14 Service Maintenance 15 Cleaning 15 Service 15 Service Items Accessories 15 Equipment Symbols 16 Specifications 18 Electromagnetic Emissions 19 Electromagnetic Immunity 19 Recommended Separation Distances 22 ER900 and ER900 AF Event Monitors 2 This page...

Страница 4: ...symptoms Once data is recorded patients transmit the recorded ECG data over the telephone or directly to a host PC for review by a licensed physician Precautions A Patient leads must be removed from electrodes before defibrillation B Observe local laws for disposal of alkaline batteries C Do not leave the batteries in the recorder when it is not in use Damage from corrosion could result D Patient ...

Страница 5: ...ARRHYTH switch turns Brady Tachy Pause and Atrial Fibrillation on or off as a group E The Arrhythmia settings are pre defined as follows Brady Rate 45 BPM Tachy Rate 140 BPM Pause 3 0 Sec ER900 and ER900 AF Event Monitors 6 Monitor Preparation done by trained Technician 1 Remove the patient cable Open the battery compartment by removing the battery door Set the programming option switches for your...

Страница 6: ...nt cable Will be followed by a single beep and rising tones Go to step 3 Rising tone Insert patient cable Go to step 3 Rising tone and a phone ring Press RECORD SEND button for two seconds Any previously recorded events will be transmitted Listen for a falling tone at the end of the transmission Insert patient cable Will be followed by a single beep and rising tones Go to step 3 3 The monitor at t...

Страница 7: ...White Red ER900 and ER900 AF Event Monitors 10 Patient Operating Instructions The ER900 or ER900 AF monitor should be ready when you receive it from the technician If there are any problems refer to the Troubleshooting section To Hookup 1 Snap lead wires onto electrodes first then apply electrodes according to physician instructions 2 Insert the patient cable into the recorder A The monitor will m...

Страница 8: ...tones while the recording is being sent C A falling tone will sound when the transmission is complete D Pressing RECORD SEND for two seconds during transmission will stop the transmission Pressing RECORD SEND again will re transmit the recording 5 When instructed that the events have been sent successfully it is OK to erase events by reinserting patient cable You will then hear a rising tone Erase...

Страница 9: ...an battery contacts Make sure mouthpiece of phone is directly over monitor speaker No information received by receiving center Ensure RECORD SEND button is held for two seconds Mouthpiece of phone must be close to the monitor speaker hole Check telephone connection Listen to phone line before sending event s to ensure there is no noise Have patient call back and send ECG again Noise artifact on re...

Страница 10: ...recorder to make shipping arrangements Note there isn t any preventative inspection or maintenance that can be performed by the end user Service Items Accessories Description Part Number Patient lead 1 channel shielded 350 0173 09 Monitor belt clip Holster 100 1764 002 Necklace 350 0180 00 Operator manual 600 0603 00 AAA I E C LR03 Alkaline Battery 200 2492 001 ER900 and ER900 AF Event Monitors 16...

Страница 11: ...ry Max event time 30 minutes Max total record time 30 minutes Type Flash Data retention Non volatile Physical Dimensions 3 30 x 2 12 x 71 83 8mm x 53 8mm x 18 0mm Weight with batteries 3 5 oz Enclosure Molded plastic Operating position Any orientation Electrical Input impedance 2M min CMR ratio 70dB AC signal range 3mV DC signal range 300mV Resolution 23uV 8bits Frequency response 05Hz to 40Hz Env...

Страница 12: ...to the public low voltage power supply network that supplies buildings used for domestic purposes Electromagnetic Immunity Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors are covered with synthetic material the relat...

Страница 13: ...ent Separation distance according to frequency of transmitter Rated maximum output power of transmitter W 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated u...

Страница 14: ...Braemar Inc Phone 800 328 2719 1285 Corporate Center Drive Suite 150 Fax 651 286 8630 Eagan MN 55121 USA E Mail Service BraemarInc com Copyright 2006 Braemar Inc All rights reserved ...

Отзывы: