ER900 and ER900 AF Event Monitors
11
To Send:
1.
Remove the patient cable from the monitor.
2.
Set the monitor on a flat surface with the speaker hole up.
3.
Call the receiving center.
A.
Cell phones and VOIP phones
do not
work for the transmission.
4.
Follow receiving center instructions.
A.
When instructed, place the telephone
mouthpiece
over the
monitor speaker hole.
B.
Push the RECORD/SEND button for
two seconds
. The monitor
will make audible tones while the recording is being sent.
C.
A falling tone will sound when the transmission is complete.
D.
Pressing RECORD/SEND for two seconds during transmission
will stop the transmission. Pressing RECORD/SEND again will
re-transmit the recording.
5.
When instructed that the events have been sent successfully, it is OK
to erase events by reinserting patient cable. (You will then hear a
rising tone.)
Erase Events:
1.
Complete the “To Send” section above first.
2.
Insert the patient cable. You will hear a rising tone.
3.
The monitor is now looping and ready to record new events.
ER900 and ER900 AF Event Monitors
12
This page intentionally left blank.