If you have difficulty using this product, please verify the following:
Problem
Solution
LED blinks Red
every two seconds
•
Replace the batteries.
The batteries will last between 30 to 50 nights (8-hour recordings)
depending on the batteries used (i.e., Alkaline vs. Lithium).
LED does not blink
•
Switch off the unit. Wait 10 seconds, and then switch on the unit
again.
•
Replace the batteries.
•
Contact technical support:
North America call 1-888-462-4844
Outside North America call 1.613.831.6690
When you call, have your model number, serial number, and a pen
and paper ready.
Spikes in the signal
every 10 seconds
•
Replace the batteries.
Poor signals
•
Verify that the Ultima Dual Airflow Pressure Sensor™ is properly
attached to both the patient and the headbox.
•
Verify that all amplifier recorder connections are functional.
•