On the right side of the Ultima Dual Airflow Pressure Sensor™, there are four, 20-turn screws that 
allow you to adjust both the gain and baseline settings for both the nasal and oral outputs. The 
baseline is set to zero volts in our factory. Under most circumstances, however, you will 

NOT

 need 

to make adjustments.

Gain adjustment screws

There are two gain adjustments screws: a nasal gain screw and an oral gain screw. The nasal 
gain screw allows you to adjust the gain for all the nasal outputs. Similarly, you can adjust the 
gain for all the oral outputs with the oral gain screw. 
By adjusting the gain or sensitivity, you enlarge or reduce the size of the waveform on the 
display of your recording system. You do not change the actual recorded signal.
By turning the gain screw clockwise, you increase the sensitivity to make the signal larger. By 
turning the gain screw counterclockwise, you decrease the sensitivity to make the signal 
smaller.

Baseline adjustment screws

The baseline is set to zero volts in our factory. Under most circumstances, however, you 
will 

NOT

 need to make adjustments.

If you do want to make adjustments, there are two baseline adjustments screws: a nasal 
baseline screw and an oral baseline screw. The nasal baseline screw allows you to adjust 
the baseline for all the nasal outputs. Similarly, you can adjust the baseline for all the oral 
outputs with the oral baseline screw.

When you turn the baseline adjustment screw, you change the position of the waveform 
on the display of your recording system. You move the waveform further up or further 
down on the display so you can view the signal peak-to-peak. The baseline is set to zero 
volts in the factory. 
By turning the baseline screw clockwise, you move the baseline higher; that is, increase 
the baseline voltage. By turning the baseline screw counterclockwise, you move the 
baseline lower; that is, decrease the baseline voltage.

Caution:  

Do 

NOT

 adjust the baseline without a voltmeter.

Содержание Ultima 0580A

Страница 1: ...isateur allez www braebon com Per una traduzione italiana di questa guida dell utente vada a www braebon com F r eine deutsche bersetzung dieses Benutzerf hrers gehen Sie zu www braebon com Para una t...

Страница 2: ...HQW 6FUHZV DWWHU 6WDWXV QGLFDWRU RQQHFWLQJ WKH 8OWLPD XDO LUIORZ 3UHVVXUH 6HQVRU STEP 1 RQQHFW WKH SUHVVXUH VHQVRU WR RXU VOHHS UHFRUGLQJ V VWHP STEP 2 RQQHFW WKH VDIHW ILOWHU RQ WKH FDQQXOD WR WKH SU...

Страница 3: ...ms The differential pressure system uses either custom Braebon cannulas Model 0589 0588 0582s or others or standard oxygen O2 cannulas with the Braebon safety filter The single use nasal cannula has a...

Страница 4: ...the name for the symbol and the meaning of the symbol Warning The Warning message appears in the manual before procedures or tasks that must be strictly observed to avoid patient injury or harm Cautio...

Страница 5: ...ensor Failure to use the Braebon safety filter will void the warranty To prevent dust contamination to the pressure sensor always keep safety filters attached to the unit and change the safety filters...

Страница 6: ...irflow pressure with no snoring signal Oral Snoring Output Snoring signal from upper airway pressure vibrations Nasal Input Facilitates combining nasal and oral breathing on one channel by connecting...

Страница 7: ...al oral cannula with hydrophobic filter Model 0589 Two adult nasal micro cannulas with hydrophobic filter Model 0582s One AC Interface Cable Model 0592 One DC Interface Cable Model 0594 Two 1 5 volt A...

Страница 8: ...e of volts With the Ultima Dual Airflow Pressure Sensor you can adjust the gain and baseline controls for both the nasal and oral output channels The output range will vary according to the gain and b...

Страница 9: ...ith the nasal differential input you can subtract a known pressure from the nasal baseline by connecting your CPAP circuit to the nasal differential input The nasal differential input is your baseline...

Страница 10: ...uperimposed snoring signal Nasal Airflow Output Filtered nasal airflow pressure with no snoring signal Nasal Snoring Output Snoring signal from upper airway pressure vibrations Oral Airflow Snoring Ou...

Страница 11: ...ger By turning the gain screw counterclockwise you decrease the sensitivity to make the signal smaller Baseline adjustment screws The baseline is set to zero volts in our factory Under most circumstan...

Страница 12: ...nsor See Table 2 Battery Status Usually the batteries will last about 30 50 nights 8 hour recordings depending on the battery type used When the Battery Status Indicator LED blinks red replace your ba...

Страница 13: ...stem you are using Warning Always use a new single use Ultima Airflow Pressure Cannula Model 0589 0588 0582s or others and a new safety filter Model 0583 with each patient The safety filter is require...

Страница 14: ...r 3 seconds or longer 70 Hz or higher 20 mV cm or 1000 x increase or decrease as necessary 100 Hz or greater Airflow Output 0 05 Hz or lower 3 seconds or longer 10 Hz or higher 20 mV cm or 1000 x incr...

Страница 15: ...Sensitivity Sampling Rate Airflow Snoring Output N A 70 Hz or higher 20 mV cm or 1000 x increase or decrease as necessary 100 Hz or greater Airflow Output N A 10 Hz or higher 20 mV cm or 1000 x incre...

Страница 16: ...ter time constant High Frequency Filter Gain Sensitivity Sampling Rate Airflow Snoring Output N A 70 Hz or higher 7 N A Airflow Output N A 10 Hz or higher 7 N A Snoring Output N A 70 Hz or higher 7 N...

Страница 17: ...annula to the nasal and oral inputs Combine nasal and oral breathing into one channel Using a Braebon Ultima Oral Nasal Cannula and filter Model 0589 connect the safety filter on the cannula to the na...

Страница 18: ...ure 5 CPAP Adapter Connection and Connection to the Nasal Differential Input CPAP Figure 6 CPAP Mask Connection and CPAP Tee Connection to the Nasal Differential Input Tee Connection Tee Connection CP...

Страница 19: ...ent s ears and down the front of the chest 3 Slide the cinch tubing toward the neck for a comfortable fit under the chin 4 Mount the pressure sensor on the wall or bedside table Warning Always use a n...

Страница 20: ...gain screw counterclockwise GMXVWLQJ WKH EDVHOLQH When you turn the baseline adjustment screw you change the position of the waveform on the display of your recording system By turning the screw cloc...

Страница 21: ...ply a known pressure not exceeding 25 cm H2O to the sensor input 3 On your data acquisition system set a corresponding voltage for the known pressure 4 Repeat steps 2 and 3 for the remaining inputs On...

Страница 22: ...of contaminants between patients 4 Only clean the pressure sensor with isopropyl alcohol pads For additional information refer to the Association for Professionals in Infection Control and Epidemiolog...

Страница 23: ...s indicated at the bottom of the battery compartment 4 Slide the battery cover back onto the casing ensuring the edges are sealed Caution Use only 1 5V AA batteries or damage to the Ultima Dual Airflo...

Страница 24: ...0 seconds and then switch on the unit again Replace the batteries Contact technical support North America call 1 888 462 4844 Outside North America call 1 613 831 6690 When you call have your model nu...

Страница 25: ...filtered out Nasal Snoring Airflow filtered out Oral Airflow and Snoring Unfiltered Oral Airflow Snoring filtered out Oral Snoring Airflow filtered out Maximum Output Range Volts Gain and Baseline Va...

Страница 26: ...1RWHV 1RWHV...

Страница 27: ...1RWHV 1RWHV...

Страница 28: ...liability or otherwise In no event shall BRAEBON MEDICAL CORPORATION be liable or in any way responsible for any loss of revenues or damage direct incidental or consequential including property damag...

Отзывы: